Philips 9890 000 0268 Series Service Manual
Philips 9890 000 0268 Series Service Manual

Philips 9890 000 0268 Series Service Manual

Replacement kit rotor control hs

Advertisement

Quick Links

Replacement Kit Rotor Control HS OPTIMUS
9890 000 0268x
FILING INSTRUCTIONS
File this documentation in binder:
Subsystem manual Optimus RAD / RF / C
4512 984 28231 REV AA
0.5
© 2004 Philips Medical Systems
ALL RIGHTS RESERVED

Advertisement

Table of Contents
loading

Summary of Contents for Philips 9890 000 0268 Series

  • Page 1 Replacement Kit Rotor Control HS OPTIMUS 9890 000 0268x FILING INSTRUCTIONS File this documentation in binder: Subsystem manual Optimus RAD / RF / C 4512 984 28231 REV AA © 2004 Philips Medical Systems ALL RIGHTS RESERVED...
  • Page 2 Philips Medical Systems Replacement Kit Rotor Control HS 9890 000 0268x 4512 984 28231 REV AA...
  • Page 3 Replacement Kit Rotor Control HS SERVICE MANUAL – UNIT Replacement Kit Rotor Control HS OPTIMUS Author: G. Kramm In case there are any questions concerning this manual, please send this LOPAD via fax to 49/(0)40/5078 2481 File: Repl_kit_roco_hs_28231AA.doc List of pages and drawings (LOPAD) Manual Order No: 4512 984 28231 released: 09/2004 3...11...
  • Page 4: Table Of Contents

    4.2. Replacement procedure for OPTIMUS RAD/RF rel.2.x ............10 Adjustments................. 11 Programming ................11 6.1. Hardware ..........................11 6.2. Software............................ 11 Acceptance .................. 11 Drawings ..................11 Rk RoCo HS Optimus (04.0) © 2004 Philips Medical Systems ALL RIGHTS RESERVED...
  • Page 5: Introduction And Technical Data

    Start of installation, operation and maintenance work and especially electrical work must only be executed by trained and authorized persons. This equipment must only be serviced by properly educated service specialists who have received general and system-specific training as performed by Philips Medical Systems. Warning! The system/component must be switched OFF during replacement work.
  • Page 6: Stator Inductance Test

    SRO 25 50 serial number 89123 – 654123 Generator Optimus 50 970023 Inductance: U--V = 56; V--W = 35 Appreciate your help and many thanks in advance. Rk RoCo HS Optimus (04.0) © 2004 Philips Medical Systems ALL RIGHTS RESERVED...
  • Page 7: Preparatory Work

    Open the cover. • Turn the screws 1..2 turns counter clockwise to loosen the screws and remove the cover plate. Remove the two ground connection screws. • (04.0) Rk RoCo HS OPTIMUS © 2004 Philips Medical Systems ALL RIGHTS RESERVED...
  • Page 8 Remove the four fixing screws of the fan and turn the fan through 180° to change the airflow direction. • Initial delivery, airflow direction for Turned fan, airflow direction for Optimus rel. 3 and Optimus C generators. Optimus rel. 2 generators. Rk RoCo HS Optimus (04.0) © 2004 Philips Medical Systems ALL RIGHTS RESERVED...
  • Page 9: Replacements

    Remove EZx42/43/23 from the back panel, wait 2 minutes to have access to the generator, set mounted radiographical controller to none, no sync contact. Perform tube adaptation for all tubes. (04.0) Rk RoCo HS OPTIMUS © 2004 Philips Medical Systems ALL RIGHTS RESERVED...
  • Page 10 Due to the fact that no upgrade (or repair) of a new unit is possible, this kit is not declared as ”EU” (exchangeable unit). The old RoCol unit and cables are to be disposed of in a safe way in accordance with local safety and environmental regulations. Rk RoCo HS Optimus (04.0) © 2004 Philips Medical Systems ALL RIGHTS RESERVED...
  • Page 11: Replacement Procedure For Optimus Rad/Rf Rel.2.X

    Due to the fact that no upgrade (or repair) of a new unit is possible, this kit is not declared as ”EU” (exchangeable unit). The old RoCo unit and cables are to be disposed of in a safe way in accordance with local safety and environmental regulations. (04.0) Rk RoCo HS OPTIMUS © 2004 Philips Medical Systems ALL RIGHTS RESERVED...
  • Page 12: Adjustments

    Check jumper settings according 5Z-2. 6.2. Software Accurately described in replacement procedure. Acceptance N / A Z. Drawings Z1-13.2 Dual speed rotor control EY 5Z-2 PCB programming options Rk RoCo HS Optimus (04.0) © 2004 Philips Medical Systems ALL RIGHTS RESERVED...
  • Page 13 2 - 3 MRC solo closed 4512 113 20531 Adapter decade cable WA, 1WA, 2WA, 1WB WB/WA 102 Universal I/O interface 4512 108 1264x = 4512 113 22341 4512 113 20621 PCB programming Options (c/04.0) 5Z-2 OPTIMUS R/F Philips Medical Systems...
  • Page 14 Rotor control Coding open normal 47 m X1:8 -15V X1:4 2 - 3 X1:5 MRC Solo closed X1:6 Dual speed rotor control X1:11 9890 000 0268x EY100 X1:17 Rotor control HS 4512 108 1284. OPTIMUS R/F (d/04.1) Z1-13.2 Philips Medical Systems...

Table of Contents