G For #2001 to #2067 (machines installed with 839Y0048* or 839Y0049*)
G For #2068, #2069, or #5001 to #5004
(1) Remove the scanning optics unit.
"I Removal procedure"
(2) Write down on the notepad the reading width adjustment value indicated on the label
located on the side of the new scanning optics unit.
REFERENCE
Use this value for adjustment after installation (because it becomes inaccessible when
the scanner is pushed in).
X= .
(3) Install the new scanning optics unit.
"I Installation procedure"
CAUTION
For machines with shipment control numbers of 5004 or earlier, be sure to disconnect
the four ground cables. The machines with shipment control numbers of 5004 or
earlier do not have threaded holes for securing the ground cables. Thus, if those
cables remain connected, they cause abnormal images due to noise.
[Remove/Secure] B3x8 (x2)
Ground cable x2
(4) Perform adjustment after installation of the scanning optics unit.
"I Check/adjustment procedure"
009-058-03
08.30.2002 FM3476
Scanning optics unit
Scanning optics unit
CR-IR347 Service Manual
FR7H3742.EPS
[Remove/Secure] B3x8 (x2)
Ground cable x2
FR7H36B3.EPS
MC - 265.4
MC - 265.4