4.3.2 Deactivation of the connection ............... 20 4.4 Placement of the sensor on the patient’s body ..........21 4.5Connection of the WARMIE sensor with the WARMIE24 app in the smartphone 23 4.6 Measurement of temperature and collection of results ........23...
Page 4
4.7 User settings..................... 24 4.7.1 Enable disconnection and connection notifications ........26 4.7.2 Download measurements in the background ..........26 4.7.3 Measurement settings ................... 26 4.8 Measurement registration ................26 4.9 Measurement viewing ..................26 4.10 Temperature threshold settings ..............27 4.11 Export of measurement results ..............
This manual is an integral part of the WARMIE WM-1020 temperature monitoring system. WARMIE Sp. z o.o. shall not bear any liability nor grant any warranties in connection with the damage (also indirect damage) which occurred as a consequence of failure to observe this user manual.
1.5.X/Y 30 July 2020 1.3 Scope of application This manual applies to the WARMIE temperature monitoring system, in the version specified in table 1.1. 1.4 Symbols, markings and terms used in the manual Warnings and safety measures, non-compliance with which may result in personal harm, including serious injuries.
Page 7
/ AppStore and installed on the user’s mobile phone. The mobile phone must meet specific requirements. activation A simple activity which is aimed at putting the WARMIE Sensor into the state in which it performs measurements and transmits the results to the WARMIE Application. The sensor is delivered to the user in a non-active state and may be activated and deactivated by the user.
2 PRODUCT CHARACTERISTICS 2.1 Intended use The WARMIE medical device is a battery-powered wireless thermometer designed for the continuous measurement and monitoring of the human body tempe- rature. The measured temperature value is transmitted wirelessly to the application installed on a user’s smartphone, which enables the reading of the temperature and the registration of its changes.
Leaving the device on your body for a period longer than 30 days may lead to the development of pressure sores. Control the condition of your body at the site where the product comes into contact with the skin and change the application site maximum every 7 days.
Page 10
Other radio devices (including their antennas) should not be placed closer than 30 cm to the WARMIE sensor. The battery may only be replaced with a battery of the type indicated by the manufacturer.
Protect the device against shocks and falls. Store the product in a clean and dry place. Store the product in its original, closed packaging, out of reach and sight of children. If you notice any irregularities in the operation of the thermometer or that it is damaged, stop using the device and contact the authorised service centre indicated by the manufacturer - see section Do not repair and/or modify the product on your own.
Keep this manual for future reference, if necessary. In the case of any doubts related to the use of the product, contact the manufacturer (see section 9). The device cannot recharge the battery. 2.7 Product classification Classification Class Class according to MDD 93/42/EEC Class IIb, rule 10 Power supply Internal battery supply, replaceable battery...
Page 13
Manufacturer Before using the device, read the user manual thoroughly. Batch number The CE mark which confirms the compliance of the product with the principal requirements of EU Directives regarding the product, including the number of the Notified Body which participated in the compliance assessment process.
2.9 Product label 2.10 Environmental working, storage and transport conditions 2.10.1 Working conditions Ambient temperature from 0°C to 50°C Relative humidity from 5% to 90% RH Atmospheric pressure 700 hPa ~ 1060 hPa (corresponds to the height above (up to 2000 m above sea level) sea level) Required time of acclimatisation of the device after its application on...
3.1 Equipment The WARMIE system consists of: • the device per se - the WARMIE temperature sensor, Fig. 5a • the band (size: S/M and L) - a silicone cover is provided on a single band, Fig. 5b • the user manual •...
Android WARMIE Sp. z o.o. shall not guarantee the proper operation of the WAR- MIE24 app. on devices with an incompatible version of the operating system. The application will not operate properly, if the version of the operating system is lower...
The manufacturer will conduct tests for each newly appearing version of the system and will publish information about its compliance on the manufacturer’s website www.warmie.eu. If any non-compliance is identified, the manufacturer will try to introduce the necessary improvements and make its newer version available.
Page 18
The manufacturer shall not be held liable for the consequences of such use. Before starting the application, activate the WARMIE sensor and put it into the pairing state - this is described in section below.
Bluetooth radio connection between the WARMIE sensor and the smartpho- ne, on which the WARMIE24 app is installed, i.e. the pairing of the WARMIE sensor with the smartphone. The user performs the pairing of these two devices in a simple manner described below.
PROBE SQUEEZE THE DEVICE IN THE CEN- DIODE LIGHT TRAL PART Fig. 1: Activation of the device to start the pairing process and measurements The activation of the device will be confirmed by a short flash of the LED installed inside the casing. Note! The flash will be invisible in strong light. 4.3.2 Deactivation of the connection The connection of the smartphone with the sensor may be interrupted manually from the level of the WARMIE24 app, e.g.
The sensor has been optimised for the measurement of temperatu- re inside the armpit. Although it is possible to measure the temperature at another anatomical location using the WARMIE sensor, it must not be used to determine the patient’s body temperature.
Page 22
Placement of the sensor: The WARMIE sensor must be placed on your body using a band attached to the set. The band must not be overtightened so as to prevent the restriction of blood supply...
7 days. 4.5 Connection of the WARMIE sensor with the WARMIE24 app in the smartphone In order to establish a connection to a mobile device, an application whose installation has been described in section is required.
from section 4.4, • it is recommended that the sensor and user’s body are in the state of ther- mal balance (i.e. approx. 15 minutes have passed since the attachment of the sensor). Start the WARMIE24 application. The application begins to monitor the temperature of your body. The application receives the current results and collects them in its register.
Page 25
meters at their disposal: • Enable disconnection and connection notifications • Download the measurements in the background • Measurement settings Fig. 3: Main screen of the WARMIE24 app after it is connected with the sensor and the setting screen (upper and lower part – not visible entirely on the screen).
4.7.1 Enable disconnection and connection notifications This setting determines whether the application is to present the push notification in the case of losing the connection with the sensor, and similarly, in the case of its connection. 4.7.2 Download measurements in the background This setting determines whether the WARMIE24 application maintains the connection and downloads the measurements when the application is not active.
between the registered temperature and time. In order to access the view of the chart, press the pictogram of the chart on the main screen of the application or anywhere in the current measurement field. The temperature values displayed in the upper part of the window indicate the Highest temperature and Average temperature in the selected period (Day, 12 hours, 6 hours or 1 hour).
The breaking of a connection between the mobile with the WARMIE24 app and the WARMIE sensor does not result in the interruption of the registration pro- cess; the only result is switching to the off-line mode and a change in the interval from 1 second to 1 minute.
WARMIE sensor from the memory. Therefore, the selection of this option requires the reconnection with the sensor as is the case with the first connection. If there is a need for interruption (deactivation) of the connection be- tween the sensor and the smartphone, select ”Clear data”.
No connec- Loss of the - Information about the Possibility to disable the tion Bluetooth lack of the connection notification in the settings. connection on the main screen of It is possible to mute the with the the app. Sound push ongoing notification.
Page 31
device, near the contact surface of both halves of the casing. Gently lift the upper part of the casing in order to separate it from the lower part. The casing will be split into two halves. One part contains the round CR2032 battery, while the other part contains a printed circuit board with electronics.
You can send the e-mail message to the service centre using the follo- wing e-mail address: support@warmie.eu You can contact the manufacturer directly using the address given on p.
Enable Bluetooth and location services in the phone settings, during the first connec- tion with the device. 7.2.3 Information about the lack of connection The app may display a message that the WARMIE sensor is not connec- ted. This happens in the following cases: • the first start-up, •...
Bluetooth® 4.0, correct transmission range of up to 10 metres in free space. The actual range depends on am- bient conditions and a mobile phone’s specifications. Equipment WARMIE temperature sensor, two bands (sizes: S/M and L), silicone cover, casing opener, user manual and warranty card Power supply CR2032 lithium battery (min.
The product is designed for domestic and professional *depending on the frequency of connections of the mobile phone with the sensor, the battery life may be shortened. 9 MANUFACTURER WARMIE Sp. z o.o. warmie.eu Kasztelańska 62, 60-316 Poznań, Polska tel. +48 501 647 303...
Page 36