OPTIONAL COMPONENTS
4. As shown in the Fig. 4, pass the recorder sensor
through the sensor guide pipe from the front to the back.
The sensor guide pipe is provided on the upper left side
of the base compartment.
5. Take out the recorder sensor from the guide pipe at
the back side and pass the sensor to the chamber
through the access port. (Fig. 5)
The access port is covered with an thermal insulation
and rubber cap.
recorder sensor through it.
6. Attach the recorder sensor on the sensor cover with
the enclosed clips. Seal the access port with a silicon
and replace the recorder sensor cover. Fix the cover
to the inside wall. (Fig. 6)
7. Remove the connector cover. Connect the recorder
connector at the end of the power cord with the white
connector on the left of the base compartment. Bind
the extra lead wire of the sensor with a nylon clip on the
back of the recorder. (Fig. 7)
8. Replace the left side panel and front panel and fix
them with screws.
Remove them before passing the
Recorder
sensor
Recorder
sensor cover
connector
14
Sensor guide pipe
Fig. 4
Access port
Fig. 5
Silicon
Fig. 6
Fig. 7