Acrylic block flowmeters are available in various
sizes and ranges, with direct reading scales in both
SAE and SI units for air and water (see charts on
other side). For other gases or liquids, special scales
can be provided.
If
other than air or water, please consult chemical
compatibility data for possible effects on the meter.
These meters are manufactured of durable acrylic and
if properly installed and maintained, will provide
long-term trouble-free operation.
SPECIFICATIONS
ACCURACY:
FLOATS:
METER BODY:
FITTINGS:
ELASTOMERS:
VALVES:
MOUNTING
INSERTS:
MAXIMUM
TEMPERATURE:
MAXIMUM
PRESSURE:
UNPACKING
Precautions have been taken to prevent any damage
from occurring during shipment.
THIS FLOWMETER IS DESIGNED FOR USE WITH NON-HAZARDOUS FLUIDS AT PRESSURES UP TO 100 PSI (690 KPA) AND TEMPERATURES
UP TO 150 °f (65 °C).
HAZARIXJUS FLUIDS OR EXCEEDING THE PRESSURE AND TEMPERATURE LIMITS MAY CAUSE FAILURE WHICH COULD RESULT IN INJURY.
M3231/0918
INSTALLATION & OPERATION INSTRUCTIONS
you use this meter with fluids
.
FL-2001 to FL-2025
.
± 5% Full Scale
FL-2031 to FL-2069
± 3% Full Scale
Black Glass, Stainless Steel,
Aluminum or Black Delrin
.
Clear Acrylic
.
Brass or Stainless Steel
Buna-N with Brass
Fittings
Viton O-rings with
Stainless Steel fittings
Brass or Stainless Steel
Cartridge Tvoe (Octional)
10-32 UNF
°
150
F ( 65 °C)
100 PSIG (690 kPa)
However, if the
Do NOT USE HAZAROOUS FLUIDS AND 00 NOT EXCEED TEMPERATURE OR PRESSURE LIMITS. USE WITH
FL-2000 - FL-2069
meter is received damaged,
immediately.
the model and flow range required.
ACHIEVING ACCURATE F'LOWRATES
To obtain an accurate flowrate, the float must be read
at the position indicated on the meter. If the meter
uses a ball float, the flowrate is determined by reading
the center of the ball. Additionally, the flowmeter
should be installed in a manner, which minimizes
both external vibrations and internal flow variations.
Special care should be taken so that the connections
to the meter's inlet and outlet fittings do not overly
restrict the liquid or gas flow being metered. This
could result in a reduced flow volume, preventing the
meter
from
Furthermore, internal pressures could be affected,
which can cause inaccurate flow readings. On start­
up, slowly purge any fluid trapped in the meter.
INSTALLATION
These meters are supplied with 5/8" or 7/8" hex's on
the inlet and outlet fittings. When installing 1/8-27
MNPT or '/4-18 MNPT fittings into the meter,
the appropriate size wrench on the hex to prevent
the inlet/outlet fitting from rotating. Torque only
to 60 in-lbs.
rotate, and may damage the meter body, causing leaks
and/or meter failure.
Teflon® tape to ease installation and provide a better
seal. This meter is supplied with #10-32 threaded
inserts for mounting. When installing, use slotted
screws and torque to a maximum of 35 in-lbs.
Mounting dimensions are shown in Figure I.
CAUTION
M-3231 /0918
report it to the carrier
Before installing, verify that you have
reaching
its
maximum
Failure to do so will cause the fitting to
Use pipe thread sealant or
flowrate .
place

Advertisement

loading

Summary of Contents for Omega FL-2000 Series

  • Page 1 Flow Meters Series FL-2000, FL-2069 INSTRUCTION M3231/0918 SHEET Shop online at omega.com e-mail: info@omega.com For latest product manuals: www.omegamanual.info INSTALLATION & OPERATION INSTRUCTIONS FL-2000 - FL-2069 Acrylic block flowmeters are available in various meter is received damaged, report it to the carrier sizes and ranges, with direct reading scales in both immediately.
  • Page 2 CLEANING AN D DISAS SE MBLY RE-AS SE MBLY Occasional cleanin:s may be required if dirt appears in Check to make sure that all parts are clean and dry. the flow tube or if float movement becomes restricted. To lubricate the o-rings, apply a small amount of To clean, remove lhe top plug and remove the float.
  • Page 3 Fax: (203) 359-7700 e-mail: info@omega.com For Other Locations Visit omega.com/worldwide The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARRANTY/DISCLAIMER OMEGA ENGINEERING, INC.

This manual is also suitable for:

Fl-2069 seriesFl-2001Fl-2002Fl-2003Fl-2004Fl-2005 ... Show all