Page 2
Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho- rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service procedures.
• This Installation and Start-Up Instruction is valid for Opdimar systems: - with SUN workstation part No. 83 83 106/ Sun Blade 150 with ASW 3.1 and with serial No.W 10000 Siemens AG SPB7-230.814.01 Page 1 of 4 Opdima Medical Solutions Rev.
NOTE contains information provided with special emphasis to facilitate proper use of the equipment or proper execution of a procedure, i.e. hints, tips . Opdima SPB7-230.814.01 Page 2 of 4 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions...
Automatic exposure control (AEC) The result must be recorded in the document MAMMOMAT 300/3000, MAMMOMAT 3000 Modular incl. Stereotactic Biopsy Attachment and Opdimar Measurement Certificates. Opdima SPB7-230.814.01 Page 4 of 4 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions...
Page 9
Do not touch the pins in the contacts of the camera. The camera shall be used within 10230i C. The camera shall be transported or stored within 0240i C. Siemens AG SPB7-230.814.01 Page 1 of 2 Opdima Medical Solutions Rev.
Page 10
2 - 2 Protective measures for CCD camera This page intentionally left blank. Opdima SPB7-230.814.01 Page 2 of 2 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions...
Page 11
Use ESD-equipment, ground prior to making contact and place the components on a conductive surface. The boards contain electrostatic highly sensitive components requiring particular care in their handling. Risk of damaging components. Siemens AG SPB7-230.814.01 Page 1 of 2 Opdima Medical Solutions Rev. 01 10.04...
Page 12
3 - 2 Protective measures This page intentionally left blank. Opdima SPB7-230.814.01 Page 2 of 2 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions...
1. Only needed if a software reinstallation is necessary (see MAMMOMAT 1000/3000 Nova - Opdimar Service Instructions). Store the documentation and software CDs in a proper place for future service use. Opdima SPB7-230.814.01 Page 2 of 2 Siemens AG Installation Rev. 01 10.04 CS PS 24...
Fig. 1 MAMMOMAT 3000 Nova with digital biopsy and spot imaging system Opdimar NOTE The workstation shall be placed at least 1.5 m from the object table of the stand. Siemens AG SPB7-230.814.01 Page 1 of 24 Opdima Medical Solutions Rev.
Page 16
The holes for fastening of the biopsy controller BC and of the cable tray below the BC have been drilled in the factory. There are two cable trays of different lenghts delivered, use the suitable one. Opdima SPB7-230.814.01 Page 2 of 24 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions...
Install the three holders (2/Fig. 2) by using the nuts (3/Fig. 2). Install the two standoffs (4/Fig. 2) by using the screws (5/Fig. 2) and the washers (6/Fig. 2). 985 – 5 mm Fig. 2 Rear cover, from the back side Siemens AG SPB7-230.814.01 Page 3 of 24 Opdima Medical Solutions Rev.
Page 18
Install the rear cover (2/Fig. 3) on the stand and attach the biopsy controller (1/Fig. 3) to the holders. Switch on the power-on switch on the biopsy controller. Fig. 3 Biopsy controller Opdima SPB7-230.814.01 Page 4 of 24 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions...
6 on page 2-6, see wiring diagram doc No: SPB7-230.051.09.03.02. Perform the connection of cables according to data in the wiring diagram. Siemens AG SPB7-230.814.01 Page 5 of 24 Opdima Medical Solutions Rev.
Remove the screws (1/Fig. 5) and remove the cover (2/Fig. 5) of the object table arm. Fig. 5 Object table arm before installation Opdima SPB7-230.814.01 Page 6 of 24 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions...
Page 21
Install a new thread (1/Fig. 6). Fig. 6 Installation of thread Install the holder (1/Fig. 7) with the pull relief device, using included Allen screws (2/Fig. 7). Fig. 7 Object table arm after installation Siemens AG SPB7-230.814.01 Page 7 of 24 Opdima Medical Solutions Rev.
Attach the supplied plastic strip around the opening of the cable outlet cover (4/Fig. 8). Install the cable outlet cover (4/Fig. 8). Mount the previously removed stand covers. Opdima SPB7-230.814.01 Page 8 of 24 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions...
Connect the cable from the control unit (2/Fig. 9) for the vertically adjustment to the connector marked “Handschalter” (3/Fig. 9). Connect the cable from the motor (4/Fig. 9) for the vertically adjustment to the other connector (5/Fig. 9). Siemens AG SPB7-230.814.01 Page 9 of 24 Opdima Medical Solutions Rev.
Page 24
Fig. 10 System site overview There must be enough space around the workstation table so that NOTE it safely can be vertically adjusted. Opdima SPB7-230.814.01 Page 10 of 24 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions...
Page 25
Mount the MO unit shelf on the right below the tabletop (1/Fig. 11). Place the MO unit to the right on the shelf (1/Fig. 11). Siemens AG SPB7-230.814.01 Page 11 of 24...
-Place the plastic cover (2/Fig. 13) on the other side of the connectors. -Fasten the plastic screws (3/Fig. 13). Biopsy controller Fig. 13 Mounting the touch protection Opdima SPB7-230.814.01 Page 12 of 24 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions...
Page 27
Work- station station Power On-Off breaker Power On-Off breaker Power connection Power connection in MAMMOMAT 3000-stand in MAMMOMAT 3000-stand Fig. 14 Power distribution Siemens AG SPB7-230.814.01 Page 13 of 24 Opdima Medical Solutions Rev. 01 10.04 CS PS 24 Installation...
Page 28
MAMMOMAT power sup- ply. (The workstation table can be connected to any branch circuit.) Opdima mains power supply outlet! Fig. 15 Mains marking label Opdima SPB7-230.814.01 Page 14 of 24 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions...
Page 29
Audio module headphones connector, not used Audio module line-out connector, not used Audio module line-in connector, not used Audio module microphone connector, not used Fig. 17 Back panel overview Siemens AG SPB7-230.814.01 Page 15 of 24 Opdima Medical Solutions Rev. 01 10.04...
Page 30
5 - 16 Installation Fig. 18 Port extenders to the back panel of the workstation Opdima SPB7-230.814.01 Page 16 of 24 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions...
AC IN SCSI ID SCSI CONNECTOR F.GND MO AC Biopsy power controller Printer/ Camera WS Mains Network controller power supply connector Fig. 19 Siemens AG SPB7-230.814.01 Page 17 of 24 Opdima Medical Solutions Rev. 01 10.04 CS PS 24 Installation...
- one mouse. There is a KVM switch delivered connecting these three input/output devices alternatingly either to the WH AWS of the MAMMOMAT Novation DR or to the SUN-Blade 150 com- puter workstation of the Opdimar. If reprogramming of the KVM switch is necessary follow the description below.
Connect the monitor to the workstation with the accompanying cables to the connectors on the backside of the monitor according to Fig. 20. Sun Ultra 1, Sun Ultra 10: 13W3-4 BNC, Siemens Part. no. 66 33 684 Red cable Green cable...
Page 34
Contrast Backlight Color Position Zoom: Fill all Source: Analog RGB Auto: Auto brightness: Auto position zoom: Auto frequency phase: Auto all Language: English Opdima SPB7-230.814.01 Page 20 of 24 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions...
Connect one end of the network cable (included in the delivery) to the workstation. Connect the other end of the network cable to a network socket. Log in with the user name “service”. The password is obtained from Siemens Service Centre.
It is also possible to connect a stand alone postscript printer to NOTE the parallel connector of the workstation, see Select printer in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions. Opdima SPB7-230.814.01 Page 22 of 24 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions...
Functional test (workstation table) Before the installed workstation table (option) is handed over to the customer, check that the workstation table can be vertically adjusted between the highest and lowest possible positions properly and safety. Siemens AG SPB7-230.814.01 Page 23 of 24 Opdima Medical Solutions Rev.
Page 38
5 - 24 Installation This page intentionally left blank. Opdima SPB7-230.814.01 Page 24 of 24 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions...
Measures after service in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions • Fill in the IVK list or report the IVK components to your responsible uptime service center (USC). Siemens AG SPB7-230.814.01 Page 1 of 4 Opdima Medical Solutions Rev.
The workstation is started according to instructions in Starting up the workstation and login in Supplement to the Instructions for Use MAMMOMAT Novation - Opdimar. Use the user name “service” and the password obtained from Siemens service center when entering the system. NOTE...
DHHS Maintenance Instructions supplied with the MAMMOMAT can be removed. However, the DHHS Measurement Certificates supplied with the MAMMOMAT should be kept for future refer- ence, if it contains previous measurement results. Siemens AG SPB7-230.814.01 Page 3 of 4 Opdima Medical Solutions Rev.
Page 42
6 - 4 Final procedures This page intentionally left blank. Opdima SPB7-230.814.01 Page 4 of 4 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions...
Page 43
Changes to previous version 7 - 1 Chapter Title Changes to previous version n.a., new version! Siemens AG SPB7-230.814.01 Page 1 of 2 Opdima Medical Solutions Rev. 01 10.04 CS PS 24 Installation...
Page 44
7 - 2 Changes to previous version This page intentionally left blank. Opdima SPB7-230.814.01 Page 2 of 2 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions...