Mechanical systems
After adjustments, make sure the brake wire link is oriented according to pos. 2 / Fig. 24.
Lock its position using the nuts (pos. 3 / Fig. 24).
1
3
Fig. 24
Adjustment for Type 2
Adjust the screw (pos 1 / Fig. 25) so that the arm can fall back just enough to release the
brake. Lock its position with the nut.
Tighten the adjusting screw with the brake released (pos. 2 / Fig. 25) until the required
brake force is obtained.
1
Fig. 25
After adjustments, make sure the brake wire link is oriented according to pos. 2 / Fig. 24.
Lock its position using the nuts (pos. 3 / Fig. 25).