Control Self Diagnostics
Control prompts will only display when a fault or alarm condition exists.
ALARM Codes "S.ERR" & "- - - -"
ALARM Code "E333"
or
Indicates Internal Circuit Error
Indicates Input Error
Check to make sure Thermocouple
Turn Controller OFF and On. If problem
wiring is connected securely
remains, replace Controller
Common Replacement Components
All replacement components are readily available and are easily replaced in the field.
COMPONENT
MODEL
VOLTAGE
PART #
Digital Controller
All
All
101-1230
Relay
All
All
401-1235
Push Button Switch
All
115V
401-2213
Push Button Switch
All
401-2213-1
230V
Thermocouple
All
All
401-1231
Magnetic Catch (set)
All
All
401-1214
10-140AE
Aluminum Shelf
All
401-6033
10-180AE
12-140AE
All
F-6043A
Aluminum Shelf
12-180AE
10-140AE
Shelf Supports (2)
All
101-1001
10-180AE
12-140AE
Shelf Supports (2)
All
101-3001
12-180AE
Maintenance / Control Calibration / Cleaning Procedures
To clean interior and exterior surfaces, use a damp cloth with or without an all-purpose cleaner. Avoid commercially available
oven cleaners. The acrylic door should only be cleaned using a lint-free cloth, with or without water. Paper towels can mar the
surface of the acrylic door. Use of any commercial cleansers on the acrylic door will cause crazing and cracking of the surface
of the acrylic over time. Periodically, check the accuracy of the control's temperature display against a known accurate or
calibrated device. This should be done with an empty chamber after the set temperature becomes steady (typically after 45 - 60
minutes). Calibrate the control in the control's function menu, level 3 (see page 3).
• If liquid is spilled inside the unit, disconnect it from the power supply and have it checked by a competent person.
• It is the user's responsability to carry out the appropriate decontamination if hazardous material is spilled on or inside the unit.
• It is the user's responsability not to use decontamination or cleaning agents that could cause a hazard as a result of a reaction
with parts of the equipment or with material contained in it.
If unsure of the compatibility of decontamination or cleaning agents consult the manufacturer.
Technical Support
If you have any questions or need technical assistance, please contact Quincy Lab Tech Support at:
Email: information@quincylab.com
Voice: 800-482-HEAT (4328)
Fax: 773-622-2282
Limited Warranty
Quincy Lab, Inc. warrants to the original purchaser that this product will be free from defects in material and workmanship
under normal use throughout the warranty period. The standard warranty period for this
instrument is eighteen months from date of shipment. The instrument warranty is supplemented with a
three year warranty on the heating element. Please refer to your invoice or shipping documents to
determine the active warranty period. This warranty covers parts & labor (labor at factory only) and
shipping cost for replacement parts.
ALARM Code "E111"
Indicates Internal Memory Error
Turn Controller OFF and On. If problem
remains, replace Controller
COMPONENT
MODEL
VOLTAGE
PART #
Acrylic Door/Hinge
10-140AE
All
401-1010
12-140AE
All
401-1211
Acrylic Door/Hinge
115V
401-2025
Motor w/ Fan & Brkt
All
Motor w/ Fan & Brkt
All
230V
401-2025-1
FUSE (3 AMP)
10-140AE-1
230V
Q-1196
250V FA 5 X 20 mm
12-140AE-1
FUSE (5 AMP)
10-180AE-1
230V
Q-1193
250V FA 5 X 20 mm
12-180AE-1
Fuse Holder (Black)
All
All
Q-1197
Fuse Holder (Red)
All
All
Q-1198
6' Cord & Plug
All
115V
101-1803
6' Cord & Plug
All
230V
101-1603-1
Quincy Lab, Inc.
1925 N Leamington Ave
Chicago, Illinois 60639
Quincy Lab, Inc.
Model 12-140AE-1
230 Volt
SPECIFICATIONS
Interior Dimensions
INCHES W x H x D
Exterior Dimensions
INCHES W x H x D
Weight (lbs)
Cubic Foot Capacity
Standard Electrical
VOLTS / WATTS
FREQUENCY (Hz)
Temperature Range
Ambient +8 C to:
Common Unit Specifications
Operating Environment: Indoor use, altitude to 6,500 ft. (2,000m) Installation Category II,
Storage Temperature:
Approvals:
Compliance:
Common Unit Construction
Exterior: Powder-Coated Steel
Insulation: Fiberglass
Thermo-control: PID Microprocessor
Issue Rev A 0519
Copyright Quincy Lab Inc. 2019
Quincy Lab, Inc.
PAGE 4
Model Series 140AE-1 & 180AE-1 (230 Volt)
General Purpose Incubators
OPERATING MANUAL
MODEL
MODEL
12-140AE-1
10-140AE-1
12x10x10
16x15.5x12
31x25x25
46x41x30
(CM) W x H x D
13x15x11
19x21x13
33x38x28
48x53x33
(CM) W x H x D
23 lbs
39 lbs
230 / 200
230 / 295
50/ 60
50/ 60
o
62°C
62°C
Pollution Degree 2, ambient temperature 10
80% RH maximum.
o
o
o
o
-10
C/14
F to 70
C/158
F, 70% RH maximum.
Underwriter's Laboratory Listed, Laboratory Equipment, C/UL
United States/Canadian. E212550 (115VAC models only)
UL Standard 61010-1, IEC 61010-1, 2nd Edition.
Interior: Aluminum
Door: 140AE-1: Acrylic, 180AE-1: Steel Insulated
Heater: Resistive-Tubular Incoloy
1925 N. Leamington Ave.
Chicago, Illinois 60639
Model 10-180AE-1
230 Volt
MODEL
MODEL
10-180AE-1
12-180AE-1
12x10x10
16x15.5x12
31x25x25
46x41x30
13x15x11
19x21x11
33x38x28
48x53x33
23 lbs
39 lbs
230 / 365
230 / 465
50/ 60
50/ 60
94°C
94°C
o
C/50
o
F to 35
o
C/95
o
F,
1-800-482-HEAT (4328)
PAGE 1

Advertisement

loading

Summary of Contents for Quincy lab 140AE-1

  • Page 1 UL Standard 61010-1, IEC 61010-1, 2nd Edition. Common Unit Construction Quincy Lab, Inc. warrants to the original purchaser that this product will be free from defects in material and workmanship under normal use throughout the warranty period. The standard warranty period for this...
  • Page 2 Menu Level Functions Guide Safety Precautions Read Operating Instructions Thoroughly Prior to Operation To access menu for common menu functions, please refer to the easy-to-use Menu Guide below: Menu setting changes are Read Operating Instructions thoroughly prior to operation and observe the following safety precautions: •...

This manual is also suitable for:

180ae-110-140ae-112-140ae-110-180ae-112-180ae-1