Page 1
HI9810312 Wireless pH Tester for Beer with built-in specialized electrode Foodcare...
Page 2
Dear Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using this instrument. Customer, This manual will provide you with the necessary information for correct use of this instrument, as well as a precise idea of its versatility.
Up to three points or four points * Calibration Automatic buffer recognition with Standard buffers Hanna (pH 1.68 *, 4.01, 7.01, 10.01) or NIST (pH 1.68 *, 4.01, 6.86, 9.18) Temperature compensation Automatic (ATC) or Manual (MTC) * Body material...
Hanna Instruments HALO2 family. • The integrated Bluetooth module allows the tester to be connected to a compatible smart device with the Hanna Lab App. • The tester has a compact, waterproof casing, and automatic pH calibration at up to three points, or four points when used with the Hanna Lab App.
3. Place the new battery with positive (+) sign facing out. 4. Reset the date and time in setup, or connect to the Hanna Lab App to update it automatically. 5. Align the mark on the cover with the open lock icon ( ) on the case. Rotate the cover clockwise until the mark on the cover aligns with the closed lock icon ( ).
PAIR 8. HANNA LAB APPLICATION ® • The Hanna Lab App is available on the App Store and on Google Play. • Consult the Help section of the application for information on calibration, measurement, data logging, and sharing. • When the tester is in discoverable mode, it will appear in the list of “Available Devices”.
” is displayed briefly and the tester will return to measurement CAL SAVE mode automatically. With Hanna Lab App (Up to Four-Point Calibration) Connect the tester to the Hanna Lab App and follow the calibration procedure. See the App Help section for the calibration procedure.
10. CARE & MAINTENANCE • Never immerse the tester over the maximum immersion level. • Fresh buffers should be used for each calibration. Once the sachets are opened the buffer value can change over time. • If the electrode is slow or sluggish, soak it in cleaning solution for 20 minutes. Rinse with water and hydrate the electrode in storage solution for a minimum of 30 minutes before calibrating.
Instruments office. If under warranty, report the model number, date of purchase, serial number, and the nature of the problem. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the tester is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization (RGA) number from the Technical Service department and then send it with shipping costs prepaid.
South Korea (KCC) R-CRM-Rgd-BMD-300 Brazil (ANATEL): Contains ANATEL approved module # 00820-21-05903. Mexico (IFETEL): Este equipo contiene el módulo con IFT #: NYCE/CT/0146/17/TS. Hanna Instruments reserves the right to modify the design, construction or appearance of its products without advance notice.
Page 12
World Headquarters Hanna Instruments Inc. Highland Industrial Park 584 Park East Drive Woonsocket, RI 02895 USA www.hannainst.com MAN9810312 Printed in ROMANIA...