Interface

With Miele cleaning machines, cleaning processes can be
documented. To enable this, these cleaning machines are equipped
with a module slot on the back to take a Miele communication
module. The communication module is available from Miele. and
comes with its own operating instructions.
 Unauthorised access poses a health risk.
Settings in the cleaning machine, e.g. parameters for disinfection or
dispensing process chemicals, may be changed as a result of
unauthorised access via the network.
The cleaning machine should be operated in a separate network
segment that is physically disconnected from other network
segments, or access to the network should be restricted using a
firewall or a router configured to provide protection against
unauthorised access.
Use strong passwords to protect access to the network.
Limit access to the network to persons requiring access in the
course of their work.
Only use terminal devices (PC, printers, etc.) which comply with
EN/IEC 60950.
Contact Miele for more information about communication modules,
software and suitable printers.
The XKM 3000 L Med communication module enables the
establishment of an Ethernet interface for digital archiving of process
data via external software.
The module can be connected to a WiFi network via an existing
wireless access point.
A XKM RS232 10 Med communication module is required for direct
connection to a protocol printer.
The XKM RS232 10 Med module can also be used for connection to
a terminal or terminal emulator. The data are transmitted in ASCII