Download Print this page
GE Druck PV210 Operating Instructions
GE Druck PV210 Operating Instructions

GE Druck PV210 Operating Instructions

Low pressure and vacuum hand pump

Advertisement

Quick Links

GE Industrial
Sensing
KEY
1. Pressure release valve
2. Fine adjust vernier
3. Pressure/Vacuum selector
4. Quick-fit connectors
5. Flexible nylon hose
6. 1/4" female NPT adapter
7. Pump handle
8. Adjustable pressure relief valve
SPECIFICATION
Pressure:
0 to 1250 in. H
Vacuum:
0 to -27 in. Hg
Relief valve:
Adjustable from 20 in. H
pressure
Materials:
Bright nickel plated brass, anodized
aluminum, phosphor bronze, nitrile seals,
nylon hose
Dimensions:
6.7" x 1.8" diameter
Weight:
1 lb
DESCRIPTION
This system is a portable source of pressure and vacuum. Each pump includes a pressure/vacuum selector, a volume
control for fine adjustment and an adjustable pressure relief valve to prevent damage to sensitive instruments.
Note: Because it has a small displacement, only use the system to pressurize small volumes.
HOSE/ADAPTERS
To attach a hose (5) and adapter (6) to a connector (4), turn the knurled nut on the connector counterclockwise.
PRESSURE RELEASE VALVE (1)
Use this to reduce or release the pressure in the system. The amount of turn sets the rate to
release the pressure. Only minimum force is necessary to seal the system.
FINE ADJUST VERNIER (2)
To make accurate adjustments to the pressure, turn the fine adjust vernier clockwise to
increase the pressure or counterclockwise to decrease the pressure.
IMPORTANT: To prevent damage, when the top of the pump body comes into view, DO
NOT use force to turn the fine adjust vernier (2) farther out.
PRESSURE/VACUUM SELECTOR (3)
Before you change the mode, make sure that the pressure release valve (1) is open. To change
the mode, use an applicable tool to push the selector (3) to the side specified on the label.
ADJUSTABLE PRESSURE RELIEF VALVE (8)
Use this to prevent damage to sensitive instruments. To adjust the maximum output pressure,
turn the screw in the bottom of the pump handle (Refer to figure).
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
99 Washington Street
Melrose, MA 02176
Fax 781-665-0780
TestEquipmentDepot.com
O, 0 to 45 psi
2
O to maximum
2
FAX 781.665.0780 - TestEquipmentDepot.com
Druck PV210 - Low
pressure hand pump
K407 Issue 1

Advertisement

loading

Summary of Contents for GE Druck PV210

  • Page 1 99 Washington Street GE Industrial Druck PV210 - Low Melrose, MA 02176 Sensing Fax 781-665-0780 pressure hand pump TestEquipmentDepot.com K407 Issue 1 1. Pressure release valve 2. Fine adjust vernier 3. Pressure/Vacuum selector 4. Quick-fit connectors 5. Flexible nylon hose 6.
  • Page 2 Operating instructions for the Druck PV210 - Low pressure hand pump GUIDELINES FOR OPERATION WARNING • DO NOT ignore the maximum operating pressure specified on the pump label. • Before you connect a pressure component to the PV210, make sure that it is isolated from the pressure supply and release the internal pressure slowly.