  • Page 1 AS100C/AS100A CPAP and Auto CPAP devices Clinical Manual...
  • Page 2 1. Entering clinical setting interface: While on standby, press both the “+” and “-” buttons and hold for three seconds. The interface will be shown as below: Figure1 Standy Figure2 Clinical setting interface A Figure3 Clinical setting interface B 2. Parameters setting From the clinical setting interface, press the “+”...
  • Page 3 From the Manual titration interface, press the Start/Stop button to start therapy. You can choose “Setting” to personalize your options, and “Message” to view therapy information. Figure4 Titration interface A Figure5 Titration interface B Service Guide 1. Entering service interface: From the therapy interface, press both the “+” and “-” buttons and hold for three seconds.
  • Page 4 Factory data reset: All settings will go back to the default value at initial state. Figure7 Default setting interface Pressure calibration(cal.): Connect the device with air resistance and pressure gage. Enter the Pressure cal. interface and set the output pressure to 20hpa. Choose “Start” to finish calibration. Figure8 Pressure calibration interface...
  • Page 5 CPAP and Auto CPAP devices Clical Manual V1.1 October, 2018 Xiaoniu health Co., Ltd.
  • Page 6 This manual No.:A030004 Directive 93/42/EEC concerning Medical Devices Xiaoniu health Co., Ltd. Tel: +8610-58411198 Fax: +8610-58411198-6300 Edition:V1.1 October 2018...