AS100C/AS100A CPAP and Auto CPAP devices Clinical Manual...
Page 2
1. Entering clinical setting interface: While on standby, press both the “+” and “-” buttons and hold for three seconds. The interface will be shown as below: Figure1 Standy Figure2 Clinical setting interface A Figure3 Clinical setting interface B 2. Parameters setting From the clinical setting interface, press the “+”...
Page 3
From the Manual titration interface, press the Start/Stop button to start therapy. You can choose “Setting” to personalize your options, and “Message” to view therapy information. Figure4 Titration interface A Figure5 Titration interface B Service Guide 1. Entering service interface: From the therapy interface, press both the “+” and “-” buttons and hold for three seconds.
Page 4
Factory data reset: All settings will go back to the default value at initial state. Figure7 Default setting interface Pressure calibration(cal.): Connect the device with air resistance and pressure gage. Enter the Pressure cal. interface and set the output pressure to 20hpa. Choose “Start” to finish calibration. Figure8 Pressure calibration interface...
Page 5
CPAP and Auto CPAP devices Clical Manual V1.1 October, 2018 Xiaoniu health Co., Ltd.
Page 6
This manual No.:A030004 Directive 93/42/EEC concerning Medical Devices Xiaoniu health Co., Ltd. Tel: +8610-58411198 Fax: +8610-58411198-6300 Edition:V1.1 October 2018...