Horizontal Or X-Axis; Main Time Base; Delayed Time Base - Philips PM 3214 Instruction Manual

25 mhz dual channel oscilloscope
Table of Contents

Advertisement

Available languages

Available languages

11
1.2.3
Horizontal or X-axis
Horizontal deflection can be obtained from either the Main time base or the Delayed time base or a combi-
nation of the two, or from the signal source selected for Xdeflection. In this case X-Y diagrams can be
displayed using A, B, the Ext. input connector, or line as a signal source for horizontal deflection.
1.2.4
1.2.5
Display modes
Main time base
Operation
Time coefficients
Continuous control range
Coefficient error
Magnification
Magnifier error
Max. effective time
coefficient
Delayed time base
Operation
Time coefficients
Continuous control range
Coefficient error
Delay time
Incremental delay time
accuracy
Delay time jitter
—Main time base
—Main time base intensified
by delayed time base
—Main time base and delayed
time base alternately displayed
— Delayed time base
— XY or XY/Y operation
Automatic
Triggered
0,5s/DIV
0,2µs/DIV
1 : > 2,5
± 3%
10x
± 2%
20 ns/DIV
Delayed time base either
starts immediately after
delay time or is triggerable
after the delay time, by the
selected delayed time base
trigger source
1 mS/DIV — 0,2 µs/DIV
1 : ~ 2,5
± 3%
Insteps variable with main
time base.
Continuously variable with
10-turn potentiometer
between 0 x and 10 x the
time coefficient of the
main time base
0,5%
n : ~ 20.000
X deflection by:
— Channel A signal
— Channel B signal
— Signal applied to EXT connector of
main time base
—Mains frequency
Possibility of automatic free-running
in the absence of triggering signals
1-2-5 sequence