Summary of Contents for Dabi Atlante Syncrus G3

  • Page 2: Presentation Of Manual

    PRESENTATION OF MANUAL INSTRUCTIONS FOR USE Technical Name: Dental Delivery Units and Accessories Trade Name: Dental Delivery Units Model: Syncrus G3 Manufacturer/ Distribuitor: Alliage S/A Indústrias Médico Odontológica C.N.P.J. 55.979.736/0001-45 - Insc. Est. 582.002.897.114 Rod. Abrão Assed, Km 53+450m - Cx. Postal 782 CEP 14097-500 Ribeirão Preto - S.P.
  • Page 3: Table Of Contents

    INDEX 02 PRESENTATION OF MANUAL 04 IDENTIFICATION OF EQUIPMENT 04 Indication of Equipment 04 Purpose of the equipment 04 Principles and essentials for the product’s functioning 05 Description of the Equipment 07 MODULES, ACCESSORIES, OPTIONS AND MATERIALS OF CONSUMPTION 12 TECHNICAL SPECIFICATIONS 12 General features 14 Electromagnetic Emissions 18 Dimensions...
  • Page 4: Identification Of Equipment

    IDENTIFICATION OF EQUIPMENT Dear Customer This manual is a general presentation of your product and it will give you important details to help you to solve possible problems. Please, read it and keep this with you. Indication of Equipment This equipment is for dental use use only. It must be operated and utilized by specialized professional (certified professional, according to the legislation of the country) and following the instructions of the manual.
  • Page 5 IDENTIFICATION OF EQUIPMENT Description of Equipment Delivery Unit, for actuation and control of the syringe, rotary instruments and others, providing the best proximity to the work area; ambidextrous (serves right and left-handed users). Structure assembly manufactured with steel with ABS injected body provided with anti-UV protection. Smooth paint high glossy epoxy-based, cured in an oven at 250°C, with phosphate treatment resistant to corrosion and cleaning materials.
  • Page 6 IDENTIFICATION OF EQUIPMENT - Shows the elapsed time and the end of the operation. - No special optical filters. - Low power consumption. - Low cost of replacement. The cold light does not emit heat as conventional bulbs - Low temperature light polymerizes the resin without damaging the tooth pulp and prevents thermal expansion problems.
  • Page 7: Modules, Accessories, Options And Materials Of Consumption

    MODULES, ACCESSORIES, OPTIONS AND MATERIALS OF CON- SUMPTION The contents of this page are of an informative nature, the equipment being able to differ from that illustrated. So, upon acquiring the product check the technical compatibilty between equipment, coupling and accessories. side view bottom view 01 - Bilateral Catcher...
  • Page 8 MODULES, ACCESSORIES, OPTIONS AND MATERIALS OF CON- SUMPTION...
  • Page 9 MODULES, ACCESSORIES, OPTIONS AND MATERIALS OF CON- SUMPTION 01 - Terminals: Borden Midwest Fiber Optic Electric micro motor 02 - Curing Light 03 - Control panel Kit Available in two versions (11/17 keys) • 04 - Triple syringe with fully hermoplastic injected body 05 - Triple syringe with fully metallic ody or with the thermoplastic injected knob 06 - Auxiliary Tray / support tools 07 - Stainless steel cover...
  • Page 10 MODULES, ACCESSORIES, OPTIONS AND MATERIALS OF CON- SUMPTION 19 - Ultrasound Kit • Available in versions with either fixed or detachable transducer with or without illumination. 20 - Kit MME Attention - The drawings (page 08 and 09) illustrates all optional items; therefore, your equipment will consist only of items selected during your purchase option.
  • Page 11 MODULES, ACCESSORIES, OPTIONS AND MATERIALS OF CON- SUMPTION The Delivery Unit may comprise: Optional Initials Borden terminal Midwest Terminal Fiber Optic Terminal Electric micro motor Terminal Curing Light OPTI Control panel Bicarbonate Jet CART coupling (C) Coupling pneumatic AIR (F) AIR (F) Coupling mechanical Ultrasound...
  • Page 12: Technical Specifications

    TECHNICAL SPECIFICATIONS General features Power Supply Voltage (coming from dental chair) 127 or 220 V~ 50/60 Hz (selectable internally) Protection against Electric Shock Type B and Class I Equipment Operation mode Continuous Input fuse (coming from dental chair) 2x10A (110/127v) or 2x5A (220v) Voltage in equipment (coming from dental chair) 12 and 24 V~ Inlet air pressure...
  • Page 13 TECHNICAL SPECIFICATIONS Wavelength 420nm - 480nm Timer 5,10,15 and 20 seconds Timer alarm Sound alarm with beep every 05 seconds and 02 beeps at the end of the cycle Activation Through the hand-piece button Light conductor Optic fiber 100% coherent that ensures the light passage without losses Weight 0,8kg Specifications of Ultrasound...
  • Page 14: Electromagnetic Emissions

    TECHNICAL SPECIFICATIONS Electromagnetic Emissions Electromagnetic Emissions Syncrus G3 Delivery Unit is made to be used in the electromagnetic environments specified Syncrus G3 Delivery Unit below. The client or the user of the must be sure that it is used in such environment.
  • Page 15 TECHNICAL SPECIFICATIONS Guidelines and manufacturer's declaration - electromagnetic immunity Syncrus G3 Delivery Unit is made to be used in the electromagnetic environments specified below. The client or the user of the Syncrus G3 Delivery Unit must be sure that it is used in such environment.
  • Page 16 TECHNICAL SPECIFICATIONS Guidelines and manufacturer's declaration - electromagnetic immunity Syncrus G3 Delivery Unit is made to be used in the electromagnetic environments specified Syncrus G3 Delivery Unit below. The client or the user of the must be sure that it is used in such environment.
  • Page 17 TECHNICAL SPECIFICATIONS Recommended distances between portable and mobile RF communication equipments and the Syncrus G3 Delivery Unit Syncrus G3 Delivery Unit is made to be used in an electromagnetic environment in which Syncrus G3 Delivery Unit RF disturbances are controlled. The client or the user of the...
  • Page 18: Dimensions

    TECHNICAL SPECIFICATIONS Dimensions (mm) Delivery Unit with CART coupling Delivery Unit with pneumatic coupling Delivery Unit with Mechanical coupling...
  • Page 19: Packing Symbols

    TECHNICAL SPECIFICATIONS Packing symbols It determines the maximum Packing to be transported and / quantity of boxes which can be or stored avoiding humidity, rains stacked during transportation and and wet floor. storage “as per packaging”. The packing must be stored and Packing to be transported and / transported away from direct sun or stored with the harrows up.
  • Page 20 TECHNICAL SPECIFICATIONS Product symbols General warning:: If the instructions are not properly General sign for mandatory action followed, the use can lead to damage to the product, user or patient High-speed with FO Curring Light Lift seat Lower seat Lift backrest Lower backrest It determines the initial position Dental Light...
  • Page 21 TECHNICAL SPECIFICATIONS Product symbols Bicarbonate Jet Ultrasound X ray view operation Flush System operation Electric low-speed-motor rotation Electric low-speed-motor inverter Emergency stop Arm fall Spit Position Activation / Back to Cup filling the last position Bowl’s water flow Triple syringe...
  • Page 22: Infra-Structure Specification

    INFRA-STRUCTURE SPECIFICATION The perfect functioning and the durability of the dental office are directly linked to the preinstallation, which must be carried out by duly trained professionals according to the instructions contained in the "Pre installation plan" which is found with the Alliage authorized technical representative. INSTALLATION OF EQUIPMENT The installation of this equipment requires the necessity of specialized technical assistance (Alliage).
  • Page 23 OPERATION OF EQUIPMENT Terminal Drive Progressive pedal (fig.1) For the operation of rotary instruments, remove support Fig.1 the instrument to be used, actuate on the foot control (b). Progressive pedal with Chip Blower / Water blocking system for hand pieces (fig.2) For the operation of rotary instruments, remove support the instrument to be used, actuate on the foot control (b).
  • Page 24 OPERATION OF EQUIPMENT High rotation terminals with double illumination system "LED selection" Select the desired illumination system through the reversal key. White light LED LED illumination with material illumination. exposer (Ultra-Vision). LED turned off Triggering through the control panel kit (PAD) 01 - Activation of the dental light 10 - Cup filling 02 - X ray view operation...
  • Page 25 OPERATION OF EQUIPMENT Activation of the dental light Press and release the key (01) to turn the Dental Light on or off. To change the luminous intensity of the Dental Light, hold down the key until the Dental Light reaches the desired intensity. Work positions The Equipo’s panel has programmable working positions.
  • Page 26 OPERATION OF EQUIPMENT Flush System Remove the reservoir uncoiling it on clockwise and make the replacement. Use a chlorinated water solution 1:500. Preparing the solution: From a solution of hypochlorite of sodium at 1%, a solution of chlorine at 500 p.p.m.
  • Page 27 OPERATION OF EQUIPMENT - Press the button to turn on the equipment (01). - Select the application mode pressing the selection button (02), which variations are: - Continuous: Maximum mode and continuous for light intensity (same luminosity from the beginning to the end of the polymerization). - Ramp: Gradual mode, light intensity increases gradually.
  • Page 28 OPERATION OF EQUIPMENT Warning This equipment shall not be used with patients suffering from respiratory or renal diseases, or under hemodialysis treatment. These cases must be surveyed by a physician. We recommend the use of safety goggles and mask when using the bicarbonate jet. Remove the upper cover (a) by unscrewing it and add a sufficient amount of sodium bicarbonate to a prophylaxis section, that is, from 20 to 40g (do not exceed the level indicated on the container).
  • Page 29 OPERATION OF EQUIPMENT Ultrasound Activation Remove the ultrasound hand piece from the holder; Choose the appropriate insert for the wanted operation according to "Techniques and Applications"; Screw the chosen insert in the hand piece with the aid of clamping key (01) and a small grip; Activate the progressive pedal (02).
  • Page 30 OPERATION OF EQUIPMENT Technical and applications All the inserts of the Ultrasound have the particularity of vibrating in an only plane (front vibrations to back, and in the axis of the insert). The lateral vibrations common to other destartarizators don’t exit, the rectilinear displacement favors more precise approach of the tooth and of the gum.
  • Page 31: Precautions, Restrictions And Warnings

    - It is not convenient that the Syncrus G3 Delivery Unit be used near to or Piled over other equipment, in case of the use near to or piled over is necessary, it is convenient that the Syncrus G3 Delivery Unit be observed to check if it is working properly in the configuration to be used.
  • Page 32: Precaution And Warnings "During The Installation" Of The Equipment

    PRECAUTIONS, RESTRICTIONS AND WARNINGS Precaution and warnings “during the installation” of the equipment - This equipment can only be unpacked and installed by an Alliage authorized technician, under the penalty of losing the warranty, as only them possess the information, the proper tools and the required training for carrying out this task.
  • Page 33: Precautions And Warnings "After" The Use Of Equipment

    PRECAUTIONS, RESTRICTIONS AND WARNINGS Precautions and warnings “after” the use of equipment - Turn off the main switch of the dental set when it is not in use for an extended period of time. - Always maintain the equipment clean for the next operation. - Do not modify any part of the equipment.
  • Page 34: Corrective And Preventive Maintenance And Preservation

    CORRECTIVE AND PREVENTIVE MAINTENANCE AND PRESERVATION Additional procedures for reuse The equipment can be reused in undetermined, i.e. unlimited, quantities, only needing to be cleaned and disinfected. Cleaning Important: In order to execute cleaning or any type of maintenance, ensure that the equipment is disconnected from the electrical network.
  • Page 35 CORRECTIVE AND PREVENTIVE MAINTENANCE AND PRESERVATION Ultrasound / Bicarbonate Jet Cleaning of terminal, transductor cover, transductor and hose: We recommend using a clean cloth, dampened with water and mild soap. Autoclavable: Transducer cover, bicarbonate jet handpiece (without the hose), Tips and Tip tightening wrench are autoclavable at a maximum temperature of 134ºC.
  • Page 36: Preventive Maintenance

    CORRECTIVE AND PREVENTIVE MAINTENANCE AND PRESERVATION Bicarbonate Jet The Delivery Unit with bicarbonate jet is doted of automatic system of depressurization and internal cleaning of the hoses and hand pieces. When we stop the actuation of the command pedal, will be a jet of air of internal swept of the whole system, however if there is blockage in the system, proceed in the following way: a) Take the hose out (01) from the beak (02), direct the end to a right place (water unit, bowl etc) and actuate the pedal to certify that the blockage is in the beak (02).
  • Page 37: Unforeseen Events – Solution Of Problems

    UNFORESEEN EVENTS – SOLUTION OF PROBLEMS Upon coming across any problem in operation, follow the instructions below to check and repair the problem, and/or get in touch with your representative. Problem Probable cause Solution Delivery Unit -Handpiece is not working. - Compressor disconnected.
  • Page 38 UNFORESEEN EVENTS – SOLUTION OF PROBLEMS Problem Probable cause Solution Ultrasound - Chair’s fuse burned. - Turn off the chair from mains -The equipment doesn’t work. power and request a Technician presence. - L a c k o f p o w e r t o t h e - Deformed insert.
  • Page 39: Equipment's Warranty

    EQUIPMENT’S WARRANTY This equipment is covered by the warranty terms and norms contained in the Warranty Certificate that accompany the product. FINAL CONSIDERATIONS Among the care you have to take with your equipment, the most important is regarding of the spare parts replacement.