GE Amersham Typhoon Operating Instructions Manual page 52

Table of Contents

Advertisement

3 System description
3.3 Stages
3.3.2 Fluor stage with accessories
Type of spacer
Sample spacer
Digitization/OD spacers
(pair)
For digitization, the spacers should be placed with the digitization label facing up, and
the arrows pointing to the scan area. For OD measurements, the spacers should be
placed with the OD label facing up.
Membrane weight
The membrane weight is used to press fluorescent membranes flat against the fluor
stage. The use of a membrane weight improves image quality, since the membrane
weight removes air bubbles and reduces vibration of the membrane during scanning.
52
Position
Placed between groove 1 and 2
Placed between groove 1 and 2, and between 3 and 4.
Amersham Typhoon Operating Instructions 29193226 AC