Operating Mode - ABB SwirlMaster FSS430 Commissioning Instructions

Vortex & swirl flowmeter
Hide thumbs Also See for SwirlMaster FSS430:
Table of Contents

Advertisement

6.4

Operating mode

The parameters for the different operating modes are described in the following table.
Operating mode /
Description
(order code)
Liquid Volume / NL1
Actual volume flow of the liquid
medium
Liquid Volume
Standard volume flow in the
(temperature
normal condition
compensated) / NL2
Liquid Mass (no
Liquid mass flow, based on direct
adjustment) / NL3
determination of the operating
density via analog input, HART
input or default setting.
Liquid Mass (density
Fluid mass flow, based on
adjustment) / NL3
reference density and density
expansion coefficient in the
normal condition
1) The highest priority of the device is to record the operating temperature.
2) The highest priority of the device is to record the density via the analog input, as long as the analog input is activated as a density input. If the analog input is not
available as a density input, the system attempts to record the density via the HART input. If both the analog input and the HART input are deactivated as a density
input, the system uses the default density value.
3) The connection via the analog input or HART input is described in the chapter titled „Electrical connections" on page 21.
32 CI/FSS/FSV430/450-EN Rev. D | SwirlMaster FSS430, FSS450 VortexMaster FSV430, FSV450
Additional parameters required
Measuring medium temperature
Reference temperature in the normal
condition
Volume expansion coefficient
2) 3)
Operating density
Measuring medium temperature
Reference temperature in the normal
condition
Density expansion coefficient
Reference density in the normal condition Device
Parameter setting
1)
With internal temperature sensor.
No information required, the measured value
from the temperature sensor is used.
Default setting for the temperature value:
Device Setup / Plant/Customized /
Compensation Setting -> Preset Int.Temp
Device
Setup
Compensation Setting -> Ref. Temperature
Device
Setup
Compensation Setting -> Volume Exp.Coef.
Via analog input:
Input/Output / Field Input / Analog In Value ->
Density
Via HART input:
Input/Output / Field Input / Hart In Value ->
Density
Default setting for the density:
Device Setup / Plant/Customized /
Compensation Setting -> Preset Density
1)
With internal temperature sensor.
No information required, the measured value
from the temperature sensor is used.
Default setting for the temperature value:
Device Setup / Plant/Customized /
Compensation Setting -> Preset Int.Temp
Device
Setup
Compensation Setting -> Ref. Temperature
Device
Setup
Compensation Setting -> Density Exp.Coef.
Setup
Compensation Setting -> Ref. Density
/
Plant/Customized
/
/
Plant/Customized
/
/
Plant/Customized
/
/
Plant/Customized
/
/
Plant/Customized
/