  • Page 1 PTB 04 ATEX 1027 API COMPLIANT Ex II 2 G Ex db IIB T6 Gb User Manual CHAMANUAL-0917 KAM CONTROLS, INC. TEL +1 713 784 0000 3939 Ann Arbor Drive FAX +1 713 784 0001 Houston, Texas 77063 USA Email Sales@Kam.com...
  • Page 2: Table Of Contents

    & temperature, and high-pressure steam refer to the Pipeline Operators’ “Health, Safety and Environmental Policy Procedures” to ensure safe installation. KAM CONTROLS, INC. reserves the right to make changes to this document without notice. KAM CONTROLS, INC. CHAMANUAL 0917...
  • Page 3: Introduction

    These beams of light pass through a 25mm window of pipeline fluids and then are reflected back through fiber optics to the electronics enclosure for a total path of 50mm. KAM uses four LED's for color determination to increase the accuracy over three-color (RGB) models.
  • Page 4: Features

    The disparity in light output versus light reflected back is then converted into an electronic or optical signature which is amplified and sent to the computer. The computer in turn utilizes a mathematical formula to determine the exact color and haze (turbidity) of the fluid in the pipeline. Final color measurement from the KAM CHA™ can be ®...
  • Page 5: Specifications

    12" to 60" – Off-the-shelf lengths are 12", 24", 36", 48", and 60" (305 mm, 610 mm, 762 mm, 914 mm, 1219 mm, 1524 mm) Pipe size: 1" to 48" (25 mm to 1219 mm) Weight: from 20 lbs. (9 kg) KAM CONTROLS, INC. CHAMANUAL 0917...
  • Page 6: Dimensional Drawings

    TABLE 2-3 DIMENSIONS INCHES Shaft Lengths are available in .5" (12.7 mm) increments. 1.25 Standard sizes are 24", 30", 36", 48", and 60". (609.6 mm, 762 mm, 914.4 7 .1 mm, 1219 mm, 1524 mm) KAM CONTROLS, INC. CHAMANUAL 0917...
  • Page 7 S P E C I F I C AT I O N S C O N T I N U E D FIG. 2-4 FLOW THROUGH CHA™ ANSI FLANGES TABLE 2-5 DIMENSIONS INCHES 11.8 7 .25 KAM CONTROLS, INC. CHAMANUAL 0917...
  • Page 8: Installation

    3 or 9 o'clock position. Fig. 3-2 The KAM CHA probe must be oriented so that the flow indicator is parallel to the flow direction of the pipe to enable flow through the measurement window.
  • Page 9: Main Line

    MAIN LINE INSTALLATION CAUTION: For probe installations and removal when FIG. 3-4 KAM CHA™ INSTALLED ON A MAIN PIPE ® line pressure is above 100 psi, a KAM IT Insertion Tool ® must be used. Locking Collar Without the IT, pressure above 100 psi could cause the...
  • Page 10: Removal

    (KAM CONTROLS recommends using thread sealant and not Teflon tape for the threaded CHA™). Open full-opening valve. Loosen Socket Cap Screws on the Locking Collar. REMINDER: Line pressure must be below 100 psi for CHA installation and removal or a KAM IT Insertion ®...
  • Page 11 Their major function is to apply light pressure on the chevron packing to ensure a seal between the Seal Housing body and the insertion shaft. FIG. 3-10 FIG. 3-11 Flow Indicator Flow Measurement Area Flow Direction KAM CONTROLS, INC. CHAMANUAL 0917...
  • Page 12 To remove the CHA™ sensor, first disconnect all electrical connections to the CHA™ enclosure. Make sure that the line pressure is below 100 psi. If line pressure is over 100 psi, install a KAM IT Insertion tool onto the unit and follow extraction directions per the IT manual.
  • Page 13: Power Wiring

    For installations in the EU: all cable entries stopping cables or plugs shall be certified per the requirements in EN60079-1. CAUTION: The KAM CHA™ is NOT loop powered. Connection of power wiring to 4-20 mA (LOOP1 and LOOP2, Fig. ®...
  • Page 14: Communications Wiring

    The RS485 serial port can be used for X,Y,Z data or MODBUS (see pg. 17 of this manual). When wiring is complete ensure that the lid is properly placed back in place and that all bolts are tightened. KAM CONTROLS, INC. CHAMANUAL 0917...
  • Page 15 FIG. 3-15 SCALE JUMPER INPUT JUMPER OUTPUT PROPER JUMPER PLACEMENT ACROSS INPUT AND OUTPUT KAM CHA 3939 Ann Arbor Drive Houston, Texas 77063, USA +1 713 784 0000 www.KAM.com KAM CONTROLS, INC. CHAMANUAL 0917...
  • Page 16: Operations And Maintenance

    A blank screen with prompt will now open. To display CHA readings type one of three prompts, E1, E2, or E3. E1: Displays values for Z, Y, Z, and T (haze) E2: Displays values for ASTM, Saybolt, PtCo (Platinum Cobalt), and T (haze) E3: Displays all values KAM CONTROLS, INC. CHAMANUAL 0917...
  • Page 17: Cleaning And Inspection

    O P E R AT I O N S A N D M A I N T E N A N C E C O N T I N U E D CLEANING AND INSPECTION Under normal operation, the KAM CHA should not require cleaning, unless pipeline usage is limited to a small number of products, or products are contaminated with debris, organics, etc. Gasoline products or jet fuel in the pipeline will clean the CHA without removal.
  • Page 18: Modbus Configuration

    32 bit Float Blue LED signal transmission percent 40113 32 bit Float Green LED signal transmission percent 40115 32 bit Float Yellow LED signal transmission percent 40117 32 bit Float Red LED signal transmission percent 40119 32 bit Float IR LED signal transmission percent 40121 32 bit Float X color coordinate 40123 32 bit Float Y color coordinate 40125 32 bit Float Z color coordinate 40127 32 bit Float Haze output signal 0-100 40129 32 bit Float ASTM 1500 color signal .5 to 8 40131 32 bit Float Saybolt color signal 30 to -16 40133 32 bit Float PT-Cobalt color signal 0-500 KAM CONTROLS, INC. CHAMANUAL 0917...