  • Page 1 User’s Guide Shop online at omega.com e-mail: info omega.com For latest product manuals: www.omegamanual.info Stainless Steel FTB700 SERIES Inline Turbine Meters...
  • Page 2 Fax: (203) 359-7700 e-mail: info@omega.com For Other Locations Visit omega.com/worldwide The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.
  • Page 3: Table Of Contents

    TABLE OF CONTENTS FTB700 SERIES INSTRUCTIONS FTB700-SERIES - INLINE TURBINE METERS General Information General Information ................................Page 4 Specifications ....................................Page 4 Electronic Options Specification ............................Page 4 Installation, Maintenance, and Repair Installation....................................Page 5 Connections....................................Page 5 Maintenance and Repair ...............................Page 5 Replacement Parts ..................................Page 6...
  • Page 4: General Information

    3” 4” 6” 8” Minimum Maximum 1200 1200 3000 * Specifications subject to change. Please consult our web site for current data (omega.com). Electronic Options Specifications* FTB700 FTB700 DPF144-P/DPF144-W Power 6-36 Vdc 7-30 Vdc 7-30 Vdc, 4mA (4-20 mA when loop powered)
  • Page 5: Installation, Maintenance, And Repair

    O-ring seal. Reverse the procedure to reinstall, after coating the O-ring with lubricant (plastic Position. The FTB700 Series are all-position meters, compatible in the PVC model). Do not overtighten. operable in a vertical or horizontal position, with the meter insert in any radial position.
  • Page 6: Replacement Parts

    INSTALLATION, MAINTENANCE AND REPAIR FTB700 SERIES INSTRUCTIONS REPLACEMENT PARTS ALL METERS White Blue Housing Housing Upper housing/electronics Contact Contact service service representative representative for your for your specific model specific model Housing gasket/seal 102025 100411 Lower housing Not field Not field...
  • Page 7: General Information

    The FTB700-T will operate on a relatively wide range of current loop voltages, 24 to 36 Vdc. Lower voltages limit the load that can be applied to the loop without distortion of the signal. (See Load/Supply chart if there is a question regarding voltage vs.
  • Page 8: Features

    4-20 mA adjustment switches Installation The upper portion of the housing must be removed to Consult the FTB700-T Connections diagram before make connections. Use a standard hex wrench (5/32” or 4 connecting to the current loop and flow sensor. Be careful mm) to loosen the screws, then remove the upper half.
  • Page 9: Settings And Calibration

    FTB700-T INSTRUCTIONS Setting Averaging Time Setting Frequency The FTB700-T converts a train of off/on pulses from the For most applications, this step can be ignored, as the flow sensor into a continuous milliAmp signal that ranges standard setting will work fine. However, when a particularly...
  • Page 10 TROUBLESHOOTING FTB700-T INSTRUCTIONS Problem Probable Causes Things to try… No analog signal at reading Break in current loop Check if loop indicator light is on device Dead power supply Check multimeter voltage on power supply Reversed polarity Check polarity Output stick at 4 mA...
  • Page 11 Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification.
  • Page 12 Where Do I Find Everything I Need for Process Measurement and Control? OMEGA…Of Course! Shop online at omega.com TEMPERATURE M U Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies M U Wire: Thermocouple, RTD & Thermistor M U Calibrators & Ice Point References M U Recorders, Controllers &...