Flame Photometric Detector
3.5 Base3 Detector Personality Module (DPM)
Detector Control Paths
Several control signals are available to control various detector functions as shown below.
Network
GCP
SYSCON
Color
Touch
Display
Figure 3-8
Maxum II Detector Control Functions
Location ID Switch
The Location ID Switch, shown previously in the photograph, selects the DPM location that is
incorporated in the address, to be reported in the results.
The DPM I
distribution board. In this scenario, the following values are applied:
NOTE:
If the DPM I
value is determined by the SNE.
26
2
2
I
C
I
C
BaseDPM
PECM
C port is connected directly to the system controller via the PECM or a wiring
2
Switch Value
1
2
3
C port is connected to an SNE, the value is always set to "1". The actual location
2
Digital
(Range
Select)
Mezzanine
Module
A/D
(signal conditioning)
Digital (Ignite) / Power (Glow Plug)
300 V Bias (FID)
Enable (FPD)
5 V Power (FPD)
Service Manual, May 2018, A5E42019847001
FID
FPD
Analog
Output
Location
Left
Center
Right
Maxum Edition II Detectors