Configure Control - ABB AnalyzeIT AX460 User Manual

Single input ph/redox (orp) analyzer
Hide thumbs Also See for AnalyzeIT AX460:
Table of Contents

Advertisement

...4.3 Configure Control

4.3.1 Configure Single PID Controller
Controller
set to
PID
----
PID Controller
Rev.
----
Direct
Control Action
100. 0
% Man
Proportional Band
100
Secs
Integral Time
10. 0
Secs
Derivative Time
Pulse
Analog
----
Time
Output Type
See Section 4.4.
Power Recovery
Control Action
Set the required control action:
Rev.
– Reverse acting – see Fig. 1.2.
Direct – Direct acting – see Fig. 1.3.
Proportional Band
Set the required proportional band between 0.1 and 999.9% in 0.1% increments.
Integral Action Time
Set the integral action time between 1 and 7200 seconds in 1 second increments.
Set to 0 to disable integral action time.
Derivative Action Time
Set the derivative action time between 0.1 and 999.9 seconds in 0.1 second increments.
Set to 0 to disable derivative action time.
Note. Derivative control in pH applications could cause instability. It is therefore
recommended that this is set to 0 in pH applications.
Output Type
Set the required output type:
Time
– Time proportioning (relay 1)
Analog – Analog output (analog output 1)
Pulse
– Pulse frequency (relay 1)
set to
Output Type
Pulses/Minute
set to
Output Type
Cycle Time
set to
Output Type
Output Range
– continued on next page.
Time
– continued on next page.
Analog
– continued on next page.
Pulse
PROGRAMMING...
13