Optional Onboard Display With User Interface; Emi/Rfi; Emc - GE DewPro MMY31 Installation And Operation Manual

General eastern trace moisture transmitter
Hide thumbs Also See for DewPro MMY31:
Table of Contents

Advertisement

September 2005
Optional Certifications/
Approvals
Optional Onboard
Display with User
Interface

EMI/RFI

EMC

5-2
FM IS Cl. I, II, III, Div. 1, Grps. A-G, T5
FM XP-IS Cl. I, Div. 1, Grps. A-D, T5
FM NI Cl. I., Div. 2, Grps. A-D, T4A
DIP Cl. II, III, Div. 1, Grps. E-G, T5
ATEX II 3G EEx nA IIC T4
The optional onboard display with user interface uses a matrix
configurator for:
range changes
unit of measure selection
current loop adjustment
error diagnostics
current value selection for fault conditions
entering a pressure constant for ppm
Performance Criterion A:
1. Conducted Emission Test as per CISPR 11 Class A, 2004
2. Radiated Emission Test as per CISPR 11 Class A, 2004
3. Radiated Susceptibility Test as per IEC 61000-4-3, 2002
4. Electrostatic Discharge Test as per IEC 61000-4-2, 2001
5. Electrical Fast Transient Test as per IEC 61000-4-4, 2004
6. High Energy Surge Immunity Test as per IEC 61000-4-5, 2001
7. Power Frequency Magnetic Field Test as per IEC 61000-4-8, 2001
IEC 61326, Industrial Locations
.
v