Chapter 2 Command Summary
Front Panel to SCPI Command Maps
Input Channels Conditioning Keys to SCPI Command Map
(Cont.)
1
a.
[:SENSe]:EVENt[1|2]:LEVel[:ABSolute]:AUTO ON|OFF
b 1 . [:SENSe]:EVENt[1|2]:LEVel[:ABSolute] <numeric_value> [V]
b 2 . [:SENSe]:EVENt[1|2]:LEVel:RELative <numeric_value> [PCT]
c.
[:SENSe]:EVENt[1|2]:SLOPe POSitive | NEGative
d 1 . [:SENSe]:EVENt[1|2]:HYSTeresis:RELative 100 *
d 2 . [:SENSe]:EVENt[1|2]:HYSTeresis:RELative 50
d 3 . [:SENSe]:EVENt[1|2]:HYSTeresis:RELative 0 **
e 1 . [:SENSe]:EVENt2:FEED " [:]INPut[1]"
e 2 . [:SENSe]:EVENt2:FEED " [:]INPut2"
2
:INPut[1|2]:IMPedance <nume ric_value> [OHM]
3
:INPut[1|2]:COUPling AC|DC
4
a 1 . :INPut[1|2]:ATTenuation 1
a 2 . :INPut[1|2]:ATTenuation 10
5
:INPut[1|2]:FILTer ON | OFF
6
:INPut3:COUPling?
:INPut3:IMPedance?
__________________________
*
Note, in firmware revisions 3317, 3335 and 3402, use 0.
**
Note, in firmware revisions 3317, 3335 and 3402, use 100.
Figure 2-1. Input Channels Conditioning Keys to SCPI Command
Map (Part 2 of 2)
Programming Guide
2-5