HP R85026A Manual page 22

Detector 26.5 to 40 ghz
Table of Contents

Advertisement

RETURN LOSS
Specification
26.5 to 40 GHz; >12 dB
At +20 to +30®C(+36°to +118°F)
Two HP R85026A detectors and two waveguide directional couplers comprise a reflectometer test setup
for a return loss measurement. The test setup is calibrated using a short to minimize frequency response
errors.
The measured return loss should be greater than or equal to the specified value listed above. There is
an uncertainty range of ±0.6 dB due to coupler directivity and mismatch uncertainty.
POS Z BUNKING
DEVICE
UNDER
TEST
Figure 3-1.
Return Loss Performance Test Setup
3-2
Peifomiance Tests