Input Resistance - HP 54520 Series Service Manual

Hide thumbs Also See for 54520 Series:
Table of Contents

Advertisement

Performance Test Record
|G
HEWLETT
PACKARD
HP 54520A/54522A/54540A/54542A Oscilloscopes
Model Number
Tested by.
Serial Number
Work Order No.
Recommended Test Interval - 1 Year/2000 hours
Date
Recommended next testing,
Temperature
Test
Limits
Results
de Calibrator
5.000 V
4.990 to 5.010 Vde
Amplitude
Channel 1
Channel 2
Channel 3
Chan 4/Ext
Input
502
49.5Q to 50.52
Resistance = 1 MQ
990 kQ to 1.010 MQ
Voltage
Range
Supply
Limits
Measurement
Accuracy
5V/div
30V
29.34 V to 30.66 V
5V/div
= 15V
14.34 V to 15.66 V
5Vidiv
= 5V
4.34 V to 5.66 V
200mV-—
«1.2
1.1736 V to 1.2264 V
200mV
= 600 mV
573.6 V to 626.4V
200mV
= 300 mV_—s:
173.6 V to 226.4V
10mV
60 mV
58.68 mV to 61.32 mV
10 mV
30 mV
28.68 mV to 31.32 mV
10 mV
10 mV
8.68 mV to 11.32 mV
7mvV
42 mV
41.076 mV to 42.924 mV
7mV
21 mV
20.076 mV to 21.924 mV
7TmV
7mV
6.076 mV to 7.924 mV
Offset
Range
Offset
limits
Accuracy
200 mV/div
2.0V
1.943 to 2.057 V
100 mV/div
1.0V
0.9715 to 1.0285 V
50 mV/div
500 mV
485.75 to 514.25 mV

Advertisement

Table of Contents
loading

This manual is also suitable for:

54540 series54520a54522a54540a54542a

Table of Contents