HP 5334B Service Manual page 28

Universal counter
Hide thumbs Also See for 5334B:
Table of Contents

Advertisement

t
HP8660A/C/D
SIGNAL GENERATOR
REF IN
5)
0
50n FEEDTHROUGH
J
I
TSM-L2
1
Figure
4-4.
Channel
A
Frequency and Sensitivity Setup,
80 MHz
-
ZOO
MHz
Equipment:
...................................................................................................
Signal Generator
HP 8660AlC
Procedure:
1. Set the signal generator as follows:
Frequency
...............................................................................................................
80 MHz
Amplitude
.............................................................................................................
35 mV rms
2. Set the HP 5334B as follows:
Reinitialize the 5334B.
FUNCTION
...........................................................................................................
FREQ A
.......................................................................................................................
SENS
ON
CHAN A TRIGISENS control
...........................................................................
fully CW
3.
Connect the signal generator to the HP 5334B Input
A
using a 500 feedthrough connector as shown in
Figure
4-4.
VERIFY: The Counter displays 80 MHz
k
2 Hz.
4. Record the measurement on the Performance Test Record Card, line 13.
5. Set the signal generator as follows:
...............................................................................................................
Frequency
100 MHz
..............................................................................................................
Amplitude
35 mV rms
VERIFY: The Counter displays 100 MHz
+
2 Hz.
HP 5334B
-
Service Manual
4-18