Case Differentiations - Siemens 7MB3000 Series Operating Instructions Manual

Wall-mounted device
Table of Contents

Advertisement

5.5.3

Case differentiations

Table 5-5
Case
PM
New
New
Data synchronization: PM parameters are transferred to the AMx (x=1 or 2), parameters of the AMx are
1)
transferred to the PM
Table 5-6
Case
PM
Un‐
change
d
Un‐
change
d
Un‐
change
d
Un‐
change
d
Un‐
change
d
Data synchronization: PM parameters are transferred to the AMx (x=1 or 2), parameters of the AMx are
1)
transferred to the PM
Wall-mounted device
Operating Instructions, 07/2023, A5E31930403-AB
Plug & Measure - Commissioning
Description
AMx
New
Commissioning, PM without serial num‐
ber
New
Commissioning, each module has a seri‐
al number stored at the factory.
None of the modules has been connec‐
ted to another module before
Plug & Measure installation of an AM into an existing (configured) device / replacement
of an AM in an existing (configured) device
Description
AMx
Re‐
The AM is to be replaced by a module of
placed
the same type. The replacement AM has
a different serial number than the AM
connected previously.
Replacement of the AM electronics, only
permitted without own serial number
stored in the AM electronics
Re‐
AM electronics installed as "assigned
placed
spare part".
Re‐
An AM is replaced by an AM of a differ‐
placed
ent type.
New
An additional AM of the same type is to
be installed.
An additional AM of a previously instal‐
led type is installed again.
New
An additional AM of a type that has not
been connected before is to be installed.
Combined operation
5.5 Plug & measure permitted activities
Data synchronization/Action
Automatic data synchronization
Automatic data synchronization
Data synchronization/Action
User action. Confirm <ENTER> or reject
<ESC>.
The AM data is transmitted from the PM
to the AM after confirmation, except the
identification data and the calibration
parameters.
User action. Always confirm <ENTER>.
The AM data is transmitted from the PM
to the AM after confirmation, including
the identification data and the calibra‐
tion parameters.
Normal startup without message. The
default data of the AM set in the factory
is applied.
Automatic data synchronization
action required by user.
User action. Confirm <ENTER> or reject
<ESC>.
The data is transmitted after confirma‐
tion, except the identification data and
the calibration parameters.
Automatic data synchronization
action required by user.
.
 1)
.
 1)
, no
 1)
, no
 1)
55