Sony Tektronix VX4792 User Manual page 254

Arbitrary waveform generator
Table of Contents

Advertisement

Check Autostep Mode
WAVEFORM
OUTPUT
VX4792 User Manual
This procedure checks the operation of the Autostep mode.
Electrical Characteristic Checked: Operating mode, Autostep, on page E–4.
Equipment Required: A 50 W coaxial cable and an oscilloscope.
Prerequisites: The instrument must meet the prerequisites listed on page F–3.
Procedure:
1. Install test setup and set test equipment controls (see Figure F–10):
VX4792
Figure F-10: Autostep Mode Initial Test Setup
a. Connect the oscilloscope: Connect the WAVEFORM OUTPUT through
a coaxial cable to the oscilloscope CH1 vertical input.
b. Set the oscilloscope controls:
Vertical
CH1 coupling
CH1 scale
CH1 input impedance
Horizontal
Sweep
Trigger
Source
Coupling
Slope
Level
Mode
Performance Verification
Oscilloscope
Ch 1 Input
CH1
DC
0.5 V/div
50 W
2 ms/div
CH1
DC
Positive
100 mV
Auto
F-21