NIC2-2 · Installation and Set-up Manual
We want to create a COMM field that shall display the kV. Create a new field and under field-
definition select Type COMM. According to the Instrumentarium protocol the length of the
kV value is four. This means that we have to define the length of this COMM field to four.
Select "Comm: kV" will link the kV data to this field.
Fielddefinition
Type: Comm
Row:
Comm:
These fielddefinitions will give us the "kV"-value followed by its unit at the center bottom of
the patient data window.

3.4.2.6 Protocol set-up LORAD

No set-up is made for this protocol. The data input has the correct format which means that
no translations are needed.
Creating the picture that will be printed on the film is done in the same way as usual selecting
the COMM field type. NOTE! The communication protocol has to be set before setting up
the picture, otherwise the COMM field will display NONE. Now, select which data this
COMM field should display. One COMM field for each exposure data must be created.
Note that the data sent from a LORAD stand includes also the type, e.g. "25kV".
Example
We want to create a COMM field that shall display the kV. Create a new field and under field-
definition select Type COMM. According to the LORAD protocol the length of the kV value
is four. This means that we have to define the length of this COMM field to four. Select
Comm kV will link the kV data to this field.
Fielddefinition
Type: Comm
Row:
Comm:
This fielddefinition will give us the "kV"-value included its unit at the center bottom of the pa-
tient data window.
Triacon AB Sweden
Length:
56
Column:108
kV
Font: Small
Previous
Next
Length:
56
Column:108
kV
Font: Small
Previous
Next
19
Publication no. 201101
Field:01
4
Save
Field:01
4
Save
October 2005