● Select the next peak for the detector and change the Amount attribute. Repeat for each
peak.
● Select the next detector under the Method Peaks menu and change settings for each peak.
Repeat for all detectors in the method.
Figure 4-6
● When all calibration level settings have been modified, then you can calibrate using the
currently opened chromatograms.
● You can now save the method to the analyzer.
4.13
Run the Maxum II on Process Fluid
From the front-panel display, select the desired process stream and place the Maxum II in Run.
4.14
Check for Proper Reporting of Results
See the Gas Chromatograph Portal Reference Manual
cs/document/109762016/). In the GCP Manual, read the section on:
Network Window (From the Network Window, the user can find the analyzer and double-
●
click to open the Analyzer Window for the new analyzer.)
Installation Manual
Installation Manual, 12/2019, 2000595 Rev 1
Changing Calibration Level Settings
Initial Startup Procedures
4.14 Check for Proper Reporting of Results
(https://support.industry.siemens.com/
47