Replacing boards / replacing components
Fig. 5
Remove both attachment screws for the high voltage section from the I.I. flange of the
C-arm (Fig. 5).
Remove the high voltage section from the C-arm.
Disconnect all the plugs. Dissipate any residual voltage in the cables by shorting them to
ground.
Install the new high voltage section.
Connect the control section, securing it with the screws only after you have adjusted it.
Refer to: Adjusting the I.I. mini voltage supply.
Remount the VIDEOMED DC and tighten the clamp.