Siemens MAMMOMAT Maintenance Instruction page 19

Table of Contents

Advertisement

Disable potter bucky by inhibiting compressor release at exposure end and pressing the
"Compression +" key
Position detector at first stop on patient side
Position the 7-cm-thick Phantom so to completely cover the detector.
Repeat the test for each one of the three fields.
This configuration was so chosen as to exclude all variable elements that may exist between the
X-ray source and the detector.
Carry out an exposure and then check that [r] value is within ±10 of the value from the
phantom label.
At the same time check that [o] value is not 0 (zero) or higher than 12.
If values are out of range adjust according to Service Instruction SPB7-115.840.01... .
SIE
Calculated Dose test
Refer to Service instruction SPB7-115.840.01... and check values according to previous
calibration report.
PMF Table functions and safety locks
Check that control LED and sensing micro-switches are working properly as described in the
Instruction for use SPB7-115.629.01... .
PMF Blocking exposure release
Exposure release must be blocked under the following conditions (indicated on the control
console):
- No cassette is inserted.
- Cassette was not changed after the last exposure.
- No collimation plate
SIE
Emergency STOP
Press Emergency STOP on each side of the unit, one at a time, and check that unit is
completely switched OFF
SIE
Exposure and error data recorder
Read out the last 1300 exposures/errors memory and error log file with the Service PC and
software .
SIE
Delete Exposure and error memory
Reset it by means of control panel pushbuttons 12 + 9 (+/-) all together, long beep will show
that memory clear is in progress.
SIE
Annotate exposure counters
Unit has two different exposure counters that for different reasons can differ between them.
Annotate exposure counter from technical menu and PCB 03-188
QIQ
Testing film O.D. linearity
Attach the most frequently used exposure system (e.g. grid table 18 x 24).
Load the "test cassette" with film and insert it.
Verify same conditions of previous calibration and/or maintenance
QIQ
Testing image quality
Attach the most frequently used exposure system (e.g. grid table 18 x 24).
Load the "test cassette" with film and insert it.
Position the SIB phantom over 2 cm Plexiglas or other IQ phantom (e.g.Normi 7).
Select Zero Point mode and automatic decompression.
Compress somewhat and release an exposure.
MAMMOMAT Balance
Maintenance Instructions
SPB7-115.831.01
Page 19 of 22
Rev.01
05.04
8 Miscellaneous