Troubleshooting: General Fault Isolation
A thorough visual inspection of the equipment may help identify the
problem.
Area to Inspect
connectors and cables
circuit boards
ground wires/wiring
mounting hardware
AC power outlet
The connector assembly plunger spring is calibrated at the factory. As a
result of use over a long period of time, the spring tension may become
loose and cause the connector assembly not to stay extended out of the
docking station's base. To adjust the tension on the plunger spring, see
the applied force specification identified in the
2004740-002, Revision A"
Some remote displays will automatically adjust to the docking station's
video output signal. Other remote displays may require a manual
adjustment to display the monitor data correctly.
See
"Adjust the Remote Display (Option) Video Signal"
the remote display image shows any of the following symptoms:
The image on the remote display appears to flicker.
Part of the image on the remote display cut-off.
Dash Port 2 Docking Station
2000966-138
Problems
frayed cables or other damage
bent prongs or pins
cracked housing
loose screws
excessive tension or wear
loose connection
strain reliefs out of place
moisture, dusts, or debris (top and bottom)
loose or missing components
burn damage or smell of over-heated components
loose wires or ground strap connections
faulty wiring
pinched wires
loose or missing screws or other hardware
faulty wiring of power source
"Exploded View PN
on page 6-11.