Power On-Off breaker
Fig. 23 Power distribution
Siemens-Elema AB
Solna, Sweden
Connect the MO unit SCSI cable to the workstation SCSI connector (9/Fig. 25).
Connect the MO unit cable to one of the two connectors of the MO unit. Fit the
terminator to the other connector of the MO unit.
For switch settings, SCSI ID and termination see Ultra 10/Ultra 1 MO unit in
Mammomat 1000/3000/3000 Nova - Opdima? Service Instructions.
Make sure that the SCSI cables are properly connected and fixed
to the units and that the cable is not bent too much just behind
the MO unit. Also, check that the MO unit is terminated correctly.
If disregarded, this might generate a lot of problems e.g. failure to
write images to MO disk, corrupted MO disk or corrupted hard
disk.
Connect the mouse to the keyboard.
Connect keyboard and mouse (5/Fig. 25).
Connect monitor cable to the workstation connector (11/Fig. 25).
Connect Ethernet cable (7/Fig. 25).
Mark the wall socket connected to the same branch circuit as the MAMMOMAT
(Fig. 23) with label according to Fig. 24.
The power distribution to the system is shown in Fig. 23.
1-phase
10A
L1 N PE
Power connection
in MAMMOMAT 3000-stand
Register 3
Work-
station
Off
On
Power On-Off breaker
SPB7-230.033.10
Rev. 03
08.02
2-phase
10A
L1 L2 N
PE
Power connection
in MAMMOMAT 3000-stand
Page 21 of 26
SPS-UD
5 - 21
Work-
station