Output Trimming By Hht; Output Trimming By Pc; Additional Calibration Procedures; Pv Scaling - ABB 2600T Pressure Transmitter Series User Manual

Table of Contents

Advertisement

364xS User's Manual

Output trimming by HHT

Output trimming by Hand Held Terminal is available. Please refer to the device manual for the appropriate
operations.

Output trimming by PC

Output trimming by PC is available using dedicated Software. ABB can provide the references and the
Software. Please refer to the Software instructions for the appropriate operations.

Additional calibration procedures

The 364 pressure transmitter models allow some useful procedures below described.

PV scaling

The PV scaling operation can be used to align the "zero" of the process (e.g.: the minimum tank level) with
the "zero" reading of the transmitter. A configuration tool shall be use to perform this operation through digital
communication.
After the PV scaling all the Sensor trimming operation are disabled.
There are two different ways to perform a PV scaling:
Method 1: apply to the transmitter a pressure that corresponds to the scaling value (offset) and perform the
operation using the configuration tools (Digital LCD Integral Display, Hand Held Terminal and PC).
Method 2: calculate the scaling value (offset) and apply it to the transmitter following the operation available
on the configuration tool (Hand Held Terminal and PC). With this method it is possible to perform a scaling
operation even for a value different then zero.

PV scaling by Digital LCD Integral Display

As for method 1
Apply to the transmitter a pressure that corresponds to the scaling value, wait for a stable value.
Navigate the Digital LCD Integral Display menu, select
Scaling.
To accept the operation select
Digital LCD Integral Display Keys sequence to Auto-Set Zero Scaling: 2 – 2 – 1 – OK
As for method 2
Navigate the Digital LCD Integral Display menu, select
To modify select
Edit
then scroll the desired value with the up and down keys and confirm with
the operation for the seven digits and confirm with OK.
Note: the Maximum (+ 80% of URL) and Minimum (- 80% of URL) usable values are indicated on the Display
as reference.
Digital LCD Integral Display Keys sequence to Set the PV Scaling: 2 – 2 – 2 – Edit

Set PV scaling by HHT

PV scaling by Hand Held Terminal is available. Please refer to the device manual for the appropriate
operations.

Set PV scaling by PC

PV scaling by PC is available using dedicated Software. ABB can provide the references and the Software.
Please refer to the Software instructions for the appropriate operations.

Removing PV scaling

In case the Sensor trimming operation is necessary the PV scaling (Sensor Offset) has to be removed.
OK
and wait for the end (a bargraph will indicate the working progress).
Device Config
Device Config
>
PV Scaling
>
PV Scaling
>
51
>
Auto-Set Zero
PV Scaling.
Next.
Repeat
Operation

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

364ds364ps

Table of Contents