Controls and displays

Battery charging operation

P
The MOBILETT XP Digital is switched off.
Connect the power plug and check whether the AC symbol (~) lights up on the control
panel
(4/Fig. 13 / p.
The battery charge status displays must light up. (The batteries are charging)

Exposure operation

NOTE
NOTE
Fig. 13: Battery status display
Disconnect the main power cord from the power supply and wind it completely around
the cable spool.
Switch the unit on. Wait for the initialization process to finish (acoustic signal, KV/mAs
displays visible).
Check the battery status.
Switch to manual/analog operation (A/D button on control panel).
Use the "± kV/± mAs" keys to select different values for KV and mAs.
MOBILETT XP Digital
32).
At least one of the 3 battery status LEDs must light up for the
checks in exposure operation
Otherwise, charge the batteries until this operating state is
reached (min. charging time: 2 hrs.).
If the unit has not been used for some time, error ERR13 may ap-
pear, indicating a high leakage current at the capacitor bank.
Switch the system off and back on to correct this problem.
If the error persists even after the system has been switched off
and on several times, execute Service Program 1, "Format Capac-
itor Bank," according to the system troubleshooting instructions.
SPR8-230.814.30.03.02
11.05
(3/Fig. 13 / p.
32).