Before Using The Product - Panasonic MLS-3751 Operating Instructions Manual

High-pressure steam sterilizer
Table of Contents

Advertisement

Before Using the Product

Installation
Product environment and conditions of use
Location : Indoor
Ambient temperature : 5 °C ~ 40 °C
Relative humidity : 50 % ~ 80 %
The boiling point must be set for the product according to the altitude
of the location where the product is installed. The boiling point has been
set at 100 °C at the time of shipment from the factory. Before installing
the product in a place 500 m higher than sea level, consult your dealer.
Do not install the product in the following places. Otherwise, trouble
may occur with your product.
Places exposed to direct sunlight
Places where water splashes
Inclined surfaces (It becomes difficult to detect heating without water.)
Places exposed to air containing salt, sulfur, etc.
Humid places
Dusty places
Since steam is emitted from the exhaust tank on the rear side of the product, do
not install the product in places where an electric outlet or equipment exists.
Leave a clearance of more than 15 cm behind the product and more
than 15 cm on the sides of the product. If the product is installed in
close proximity to walls, the product may overheat, resulting in trouble.
Wall surfaces may be marred by venting steam or crack under high humidity.
Anchor the product firmly by means of tip-resistant metals (2 pcs.
included with the product).
Be sure to install the product on a level rigid surface and lock all four casters.
Be sure to mount the exhaust tank on the rear side of the product
and the tank mounting brackets to the rear side of the product.
(See page 10.)
wer supply and ground connections
Po
The unit's power cord must be connected to its own dedicated and
properly wired connection outlet where no other plugs are connected.
For 120 V units :
MLS-3751L ;
The unit should be connected to an appropriate receptacle that is rated 15.8 or
greater.
For 220 V units :
The 220 V versions of the MLS-3751/MLS-3781 Autoclave are shipped without a
plug on the power cord. Use the power-plug when the products is set. The
protective grounding conductor of power supply cord must be connected to the
contacting member of an attachment plug.
Wire size
Power cord
Type
Plug
For 230 V units :
The 230 V versions of the MLS-3751L/MLS-3781L Autoclave are shipped without a
plug on the power cord. Use the power-plug which is applied IEC60309 when
the products is set. The protective grounding conductor of power supply cord
must be connected to the contacting member of an attachment plug.
rating ;
Power cord Wire size
Plug
The unit must be connected straight to ground.
The green / yellow conductor of the power cord is the grounding wire. You must
ask your dealer or a qualified electrician to do the grounding work involved.
Altitude
: 0 m ~ 3000 m
Voltage fluctuation : ± 10 % of nominal voltage
M L S - 3 7 5 1
AW G 1 4
S J T
250 VAC, 10 A or greater 250 VAC, 20 A or greater
M L S - 3 7 5 1 L
AW G 1 4
250 VAC, 10 A or greater 250 VAC, 20 A or greater
ON
M L S - 3 7 8 1
M L S - 3 7 8 1 L
More than 15 cm
Screws
(Once remove
the 8 screws at
the rear side of
the unit and
tighten them
together with the
tip-resistant
metal.)
Tip-resistant metal
9