Installing

4.1

Safety instructions for installation

For the measurement of high O
that all parts that come in contact with the sample gas are free of oil and grease and do not
contain any particles that could react with O
We highly recommend you use specially cleaned analyzer modules with the "Cleaned for
O2" option for such measurements.
4.1.1

Rack-mounted device

Protection against incorrect use of the rack-mounted device
Use the rack-mounted device only within the voltage limits specified on the nameplate.
NOTICE
Damage to the membrane keyboard and the display
Make sure when installing the rack-mounted device that the membrane keyboard and
display are not damaged.
NOTICE
Keep the rack-mounted housing closed during operation
Do not open the rack-mounted housing during operation. Only open the rack-mounted
housing during installation, when connecting up or for maintenance purposes.
NOTICE
Installation in control cabinets
Make sure there is adequate ventilation between the devices when installing in control
cabinets. Make sure that the permitted ambient temperature listed in the technical
specifications is not exceeded.
Quick Start
Compact Operating Instructions, 06/2017, A5E31805656-05
WARNING
concentrations in excess of 25%, it is absolutely necessary
2
.
2