Siemens SIREMOBIL Iso-C Installation And Setting Instructions Manual page 4

Hide thumbs Also See for SIREMOBIL Iso-C:

Advertisement

4
If a multiformat camera is not connected
Power cable
Videoprinter with 100VAC line voltage:
Connect the line wires to transformer T1, connector TI - 34 and TI - 36.
Connect the protective conductor to the protective conductor terminal stripe (ye/gn).
Videoprinter with 120VAC line voltage:
Connect the line wires to transformer T1, connector TI - 35 and TI - 36.
Connect the protective conductor to the protective conductor terminal stripe (ye/gn).
Videoprinter with 230VAC line voltage:
Connect the line wires of the video printer at terminals X2.1 and X2.2.
Connect the protective conductor at the corresponding location (yellow/green).
Route the line wires to the video printer compartment.
Hardcopy exposure control cable
Connect the control cable included with the shipment at the Memoskop "Printer"
connector.
Route the end of the cable with the 3.5 mm jack plug to the video printer compartment.
Video cable
Connect the video cable at the Memoskop "Out3" connector.
Route the video cable to the video printer compartment.
Connect the video printer. Ensure the proper termination of the video cable
(75 ohm).
Programming
For Memoskop C (with keyboard)
Switch on the system.
Press [CTRL] and [U] simultaneously on the Memoskop keyboard.

Call up the "C Documentation" sub-menu.
Perform the configuration on monitor A for SIREMOBIL Iso-C systems with one monitor.
Perform the configuration on monitor B for SIREMOBIL Iso-C systems with two monitors.
If the system is equipped with an area dose measurement device (Diamentor), call up
the "K documentation of the dose" sub-menu and configure to "yes".
End the "User menu".
SIREMOBIL Iso-C
This calls the "User Setup menu" for Memoskop.
Register 3
Installing a Video Printer
SPR2-230.033.02
Rev. 02
10.99
Page 4 of 6
TD SD 24
Medical Engineering