Performance Tests
Check Gated Mode
4-18
Electrical
Operating mode, Gated, on page 1-4.
Characteristic
Checked
Equipment
Three 50 W coaxial cables
Required
One 50 W precision termination
One function generator
One oscilloscope
Prerequisites
The AWG2021 meets the prerequisites listed on page 4-7.
1. Install test hookup and set test equipment controls:
a. Hook up the oscilloscope: Connect the AWG2021 CH1 output connector
through the coaxial cable to the CH1 vertical input connector on the
oscilloscope.
b. Hook up the function generator: Connect the function generator output
to both the AWG2021 TRIGGER INPUT and the oscilloscope CH2
input through a coaxial cable, precision termination, and a dual input
coupler (see Figure 4–6).
AWG2021
AWG2021
Function Generator
FG 5010

Figure 4-6: Gated Mode Initial Test Hookup

c. Set oscilloscope controls:
Vertical
. . . . . . . . . . . . . . . . . . . . . .
CH1, CH2 coupling
CH1, CH2 scale
CH1 input impedance
CH2 input impedance
Mode
. . . . . . . . . . . . . . . . . . . .
Oscilloscope
TDS
540
CH1, CH2
. . . . . . . . .
DC
. . . . . . . . . . . .
0.5 V/div.
50 W
. . . . . . .
. . . . . . .
1 MW
Chop
AWG2021 Service Manual