Uninterruptible Power Supply (UPS)

NOTE: This accessory to the Panda iRes Warmer product is available only in select markets. To
understand availability within your region, contact your sales representative.
The Giraffe UPS is a medical grade power supply offering uninterruptible power supply to the
Panda iRes Warmer. The UPS is designed to securely mount to a single Panda iRes Warmer via
specific mounting hardware provided in the Giraffe UPS kit, or separately in the Shelf Hardware
Mounting Kit. The UPS may serve to benefit customers experiencing power line voltage disturbances,
or Brown Outs, or temporary power outages, or Black Outs. Per the UPS Specifications, the Giraffe
UPS provides power to the Panda iRes Warmer in situations where the product is disconnected from
a wall receptacle.
For specifications and a detailed description of the installation procedure, refer to the Giraffe UPS
Installation Instructions (provided with the UPS).

Using the In-Bed Scale (optional)

The Panda Warmer can be equipped with its own in-bed scale that is operated from the control
panel screen.
Scale installation
Panda iRes Warmer
M1110737 006
Illustration
© 2007 General Electric Company, all rights reserved
Step
1
1. Remove the clear Plexiglas plate and
mattress from the bed.
2. Fit the scale weighing platform into the
recesses in the bed.
3. Route the scale cable though the gap at
the corner of the bedside panels and plug
the connector into the jack.
4. Place the clear plate and mattress on
the weighing platform. Make sure they are
properly seated on the platform.
NOTE: The scale may be
connected or disconnected at any
time.
3-17