Power Supply Fuse Replacement - HP 37718A Installation And Verification Manual

Communications performance analyzer
Hide thumbs Also See for 37718A:
Table of Contents

Advertisement

C AU T I O N
General Information
Operators Maintenance
2 Remove the adapters from the IN and OUT ports.
3 Using the blow brush with the brush removed blow through the ferrule
of the standard flexible connector and the adapter.
If the optical fibre of the fixed connector requires further cleaning this
entails disassembly of the module which should only be carried out by
suitably trained service personnel.
4 Apply some isopropyl alcohol to a piece of the cleaning paper and clean
the barrel of the adapter. Using a new piece of cleaning paper, clean
the face of the adapter. Repeat this operation, using a new piece of
cleaning paper each time.
5
Lightly press the adhesive side of the tape provided against the front
of the adapter, then remove it quickly - repeat twice. This removes any
particles of cleaning paper which may be present.
6 Replace the adapters on the flexible connector.

Power Supply Fuse Replacement

See "Fuse Replacement" on page 2-3
1-9