*RST Response

*RST Response
The IEEE 488.2 *RST command returns the instrument to a specified
state optimized for remote operation. (Use *CLS to clear the status event
registers and the SCPI error queue.)
The states of command settings affected by the *RST command are
described in Table 2-3. Table 2-4 lists command settings that are
unaffected by *RST.
Table 2-3. HP 53150A/151A/152A *RST State
Command Header
DDT
*
:DISPlay[:WINDow]:BACKground[:STATe]
:DISPlay:ENABle
:INITiate:CONTinuous
:INPut:FILTer:[:LPASs][:STATe]
[:SENSe]:AVERage:[STATe]
[:SENSe]:AVERage:[COUNt]
[:SENSe]:FILTer:FM:AUTO
[:SENSe]:CORRection:CSET:SELect
[:SENSe]:CORRection:CSET:STATe
[:SENSe]:FREQuency:OFFset
[:SENSe]:FREQuency:OFFset:STATe
[:SENSe]:FREQuency:RESolution
[:SENSe]:FREQuency:TRACking
[:SENSe]:FUNCtion:OFF
[:SENSe]:FUNCtion[:ON]
[:SENSe]:POWer:AC:REFerence
[:SENSe]:POWer:AC:REFerence:STATe
[:SENSe]:ROSCillator:SOURce
:TRIGger[:SEQuence]:HOLDoff
Programming Guide
Parameter
<arbitrary block>
<Boolean>
<Boolean>
<Boolean>
<Boolean>
<Boolean>
<numeric_value>
<Boolean>
<character_data>
<Boolean>
<numeric_value>[frequency unit>
<Boolean>
<numeric_value>[frequency unit>
<character_program_data>
<sensor_function>
<sensor_function>
<numeric_value>
<Boolean>
INTernal | EXTernal <Boolean>
<numeric_value>
State
#14INIT
ON
ON
OFF
OFF
OFF
1
ON
CORR1
OFF
0
OFF
1 Hz
SLOW
"FREQuency 1",
"POWer 2"
"FREQuency 2"
0
OFF
INTernal
0