2
Press the F or f button to move the cursor to
"Calibration," and then press the g button.
3
Select the desired calibration type by pressing the F or
f button.
To perform calibration of each blue thermal
film
1 Press the F or f button to move the cursor to "Full
Range Calibration," and then press the g button.
2 Press the F or down f button to select the tray
containing the blue thermal film for which the
calibration is to be carried out.
3 Press the EXEC button.
Calibration printing starts, and density is
automatically adjusted during printing. The result
of automatic calibration will be effective from the
next printing operation.
4 Press the MENU button.
The menu screen closes and the unit returns to the
standard screen.
To perform calibration of another blue
thermal film
Repeat step 1 to 3.
To make an automatic calibration
The unit can perform a calibration for the first film
each time you load new film sheets into the film tray
1 or film tray 2.
1 Press the F or f buttons to move the cursor to
"Auto Calibration," and then press the g button.
2 Press the F or f button to move the cursor to
"ON."
The setting under the cursor becomes effective.
Item
Setting
ON
The printer performs the calibration
automatically using the first film sheet
each time you load new film sheets
into film tray 1 or film tray 2.
OFF
The printer does not perform the
automatic calibration (retains the
factory setting).
3 Press the MENU button.
The menu screen closes and the unit returns to the
standard screen.