Chapter 9 Troubleshooting

Probe

Condition
Causes
Reduced articulation in
Various
the bending neck.
Bending neck is not
System is in a mode
articulating.
that disables
articulation:
Steer-and-Stay, freeze
frame, menu, or
desktop.
Insertion tube does not
Tube is obstructed.
freely withdraw from
the storage reel.

Power

Condition
Causes
When plugged into AC
• Power interruption.
power, system won't
• Faulty AC power
turn on. (Power switch
connection.
is on, but its green LED
• Blown AC fuses.
is not lit.)
• No AC power.
When unplugged from
• Battery disconnected
AC power, system
or incorrectly
won't turn on. (Power
connected.
switch is on, but its
• Battery low, not
green LED is not lit.)
charging, depleted, or
bad.
• Blown battery fuse.

Lamp

Condition
Causes
Lamp does not come
Various
on when you turn on
power switch. (Green
LED on the power
switch is lit.)
Actions
• Whenever using the system, minimize bends throughout the entire length of the
insertion tube, including the end nearest to the handset. Unnecessary bending can
reduce articulation.
• Redistribute the lubricant inside the insertion tube as follows. Straighten the insertion
tube to its full length, with Person A holding it near where it joins the handset (to
stabilize the probe), and Person B holding it just before the bending neck (to prevent
damaging the bending neck). Have Person B shake the entire tube. CAUTION: Do not
pull on the probe tip or the bending neck while shaking the tube. The mechanical cables
could break.
• Exit any mode the system may be in.
• Turn power switch off. Leave the system off for at least 15 seconds so that it can
complete its shutdown sequence. Then turn it back on.
Remove the funnel. Clear any obstruction or binding. For an illustration of removing the
funnel, see
"To Clean the Insertion Tube Storage Reel"
Actions
• Leave the system off for at least 15 seconds so that it can complete its shutdown
sequence. Then turn it back on.
• Check the power cable connections.
• Check the AC power source.
• Check the AC fuses. See
"To Check or Replace the AC Fuses"
• Leave the system off for at least 15 seconds so that it can complete its shutdown
sequence. Then turn it back on.
• Check battery's connections. See
on page 123.
• Check battery's charge level. See
page 122.
• Check the battery's compliance to listed specifications.
• Recharge or replace the battery. See
• Replace battery fuse. See
Actions
See
"To Determine Whether to Replace the Lamp"
GE Inspection Technologies XLG3 VideoProbe System
on page 117.
"To Remove the Battery"
"To Check the Battery's Charge Level"
"Maintaining the Battery"
"To Check or Replace the Battery (DC) Fuse"
on page 118.
on page 120.
and
"To Install a Battery"
on
on page 122.
on page 121.