Toshiba Aplio SSA-770A Service Manual page 91

Diagnostic ultrasound system
Table of Contents

Advertisement

No. 2D730-148E*O
(3)
Removing the covers
(9)
(h)
(0
0)
Remove the two caps <A>.
Remove the two screws CA> (M4) and remove the right side cover.
Remove the two screws cB> (M4) and remove the right side cover for the
peripheral units.
Remove the two caps cB>.
Remove the two screws CC=- (M4) and remove the left side cover.
Remove the two screws CD> (M4) and remove the left side cover for the
peripheral units.
Remove the four caps CC>.
Remove the four screws cE> (M4) and remove the top cover.
Remove the four screws cF> (M4) and remove the tray.
Remove the four screws cG> (M4) and remove the top front cover.
Cap CC>
Top cover
;
I
Left side cover for
the peripheral units
Screw CD>
(SBPP4-12SCR)
I
-Screw
cE>
(SBPPC12SCR)
Left side cover
Right side cover for
the oerioheral units
Screw CC>
(SBPP4-12SCR)
Top front cover
Screw cB>
(SBPP4-12SCR)
I
I*
I
.b
Screw cG>
-b-
Right side
-b *"1
cover
(SBPP4-12SCR)
II
*\**
Screw cF>
'.
Tray
(SBPP4-12SCR)
L
Cap <A>
Screw <A>
(SBPP4-12SCR)
(4)
Perform the procedures in subsection 3.4 "Removing the Monitor Unit"
IA
! CAUTION:
Two persons
are required
when removing
the monitor
unit.
I
3-74