IEC 60601-1 classification
Type of Protection
Against Electric Shock
Degree of Protection
Against Electric Shock
Degree of Protection
Against Ingress of Water
Mode of Operation
Electrical (when the heated humidifier is used with the
DreamStation Go therapy device)
Maximum recommended pressure
Water capacity
Heater plate
Inhalation pressure drop of humidifier
Exhalation pressure drop of humidifier
Humidity
Maximum delivered gas temperature
18
Class II Equipment/Internally Powered
Type BF Applied Part
Device: Drip Proof, IP22
First characteristic numeral - 2 - Protection against ingress of solid foreign objects ≥ 12.5
mm diameter.
Explanation: Protected against access to hazardous parts with a finger and protected
against solid foreign objects of 12.5 mm diameter and greater.
Second characteristic numeral - 2 - Protection against ingress of water with harmful effects
dripping (15° tilted).
Explanation: Protected against vertically falling water drops when enclosure tilted up to 15°.
Continuous
AC Power Consumption: 100–240 VAC, 50/60 Hz, 2.0-
1.0 A
20 cm H
O
2
240 ml (8 oz.) at recommended water level
Max Temperature: 68°C (154°F)
Max: 2.1 cmH
O at 60 LPM Flow
2
Max: 2.8 cmH
O at 20 LPM Flow
2
Humidity
Output: 12 mg H
min
Measured @ expected leak across therapy pressures,
15.9-35° C, 15% RH, for all compatible tubes
<43°C (109.4°F)
O/L
2