Enterprise Access Antenna System; Apexpro And Carescape Transmitters - GE ApexPro TT Technical Manual

Designed to acquire and monitor physiological data for ambulating adult and pediatric patients within a defined coverage area.
Table of Contents

Advertisement

Equipment Overview
Design
Part number
frequency
2000673-007
600 MHz
2000673-008
450 MHz

Enterprise Access antenna system

ApexPro and CARESCAPE transmitters

2-6
Antenna type
ApexPro Active Antenna
This active antenna operates within 608-614MHz. This antenna
608-614MHz
also has a bandpass filter element that rejects signals outside of
608-614MHz, except for signals in channels 36 and 38.
ApexPro Antenna Hi-Pwr
This high-power antenna operates within 420-474MHz and has
420-474 MHz
filtering for out-of-band signals. It has >15dB rejection below
320MHz.
For technical specifications, see
Refer to the Enterprise Access System Service Manual for details on the system.
CAUTION
UNINTENTIONAL RADIO FREQUENCY (RF)
INTERFERENCE—Unintentional RF interference could degrade
the reliability and performance of the wireless data link. The facility
must maintain an RF environment free from unintentional
interference.
The ApexPro and CARESCAPE transmitters send the patient's ECG data to the
ApexPro receiver system for processing. Data is then transmitted via a dedicated
Ethernet interface to the CIC Pro center for viewing. The transmitter can also send
DINAMAP PRO data to the CIC Pro center via the DinaLink cable.
Additionally, the transmitter can send the patient's SpO2 and noninvasive blood
pressure data when the interface connector ports are enabled and when the optional
oximeter and/or Accutracker DX noninvasive blood pressure monitor are connected
to it.