Settling[:Time] - HP E1351A User Manual

Fet multiplexer
Table of Contents

Advertisement

SETTling[:TIME]

Parameters
Comments
Example
SETTling[:TIME]?
Parameters
Comments
Example
Chapter 5
[ROUTe:]SETTling[:TIME] <
of a channel closing trigger, and the pulse when the FET channel is closed.
You must specify one and only one channel for each card in the instrument
which has a channel in the channel list, and the
channels on the card. The range is 1 to 32768
Parameter Name
time
channel_list
Digital Bus Handshaking: This command allows a user-defined
settling time to allow a measurement to be made before the next
channel is closed.
Related Commands:
IMMediate
Setting 16 µsec SETTling TIME
TRIGger:SOUR DBUS
SETTling:TIME 16E-6
[ROUTe:]SETTling[:TIME]? [MIN | MAX] <
current setting for a specified channel.
and maximum allowable settings.
Parameter Name
MIN | MAX
channel_list
Returned value is given in exponential notation.
Set and Query the SETTling TIME
SETT:TIME 16E-6
SETT:TIME?
HP E1351A/53A 16-Channel FET Multiplexer Command Reference
channel_list
sets the delay between receipt
>
µ
Parameter Type
Numeric
1E-6 | 32768E-6 | MIN | MAX
Numeric
TRIGger:SOURce DBUS
Triggering on the digital bus
handshake lines
Delay = l6µsec
channel_list
or
MIN
Parameter Type
MIN = 1 µsec, MAX = 32768 µsec
Numeric
Numeric
Sets delay = 16µsec.
Returns + 1.600000E-005.
time
will then apply to all
<
>
sec.
Range of Values
only one channel
,
TRIGger:SOURce
>
returns the
returns the minimum
MAX
Range of Values
only one channel