Chapter 1 Performance Tests
Test 6: Peak Volts, Channel 1 (HP 53181A Only)
Test 6: Peak Volts, Channel 1 (HP 53181A
Only)
This test verifies the Peak Volts accuracy specification of the HP 53181A
Frequency Counter.
Equipment
HP 3325B Synthesizer
HP 10100C 50Ω Feedthrough
HP 10503A BNC Cable
Counter Setup
1
Cycle the POWER key to preset the Counter.
2
Press Other Meas key until VOLT PEAKS 1 is displayed.
3
Set CHANNEL 1 and CHANNEL 2 inputs to:
50Ω/1MΩ, 50Ω
X10 Attenuate, Off
100kHz Filter, Off
1-68
Assembly-Level Service Guide